Point of care D-dimer testing in the emergency department: a bioequivalence study
- PMID: 23301220
- PMCID: PMC3535194
- DOI: 10.3343/alm.2013.33.1.34
Point of care D-dimer testing in the emergency department: a bioequivalence study
Abstract
Background: D-dimer is used widely as a diagnostic aid in low- and moderate-risk patients with suspected venous thromboembolism (VTE). While our laboratory utilizes VIDAS D-dimer analyzer (bioMérieux SA, France), our emergency department (ED) recently procured a D-dimer analyzer AQT90 FLEX (Radiometer Medical ApS, Denmark) for point of care testing (POCT) to facilitate patient management. We aimed to determine whether the time taken to receive D-dimer results using the 2 different analyzers differed significantly and to quantify the limits of agreement between the results of the 2 methods measured on the same patient.
Methods: Adult patients presenting to the ED and requiring diagnostic workup for suspected VTE were included in this prospective observational study. Patients underwent simultaneous D-dimer measurements using the 2 different analyzers.
Results: The paired results from 104 patients were analyzed. The median time for the D-dimer results from triage by VIDAS was 258 min (Inter-quartile range [IQR], 173-360) and by POCT was 146 min (IQR, 55-280.5); the median time difference was 101.5 min (IQR, 82-125.5). On an average, POCT D-dimer values were 15% lower on the same sample (limits of agreement, 34-213%). POCT predicted 83% of VIDAS positive results (sensitivity, 83.3% [95% confidence interval (CI), 70.4-91.3%]; specificity, 100% [95% CI, 93.6-100%]). All patients with positive imaging were identified correctly by both methods.
Conclusions: POCT delivers D-dimer results in significantly shorter turnaround times than pathology services; however, poor bioequivalence between VIDAS and POCT raises the issue of acceptability for use in the ED.
Keywords: Bioequivalence; D-dimer; Length of stay; Point of care systems; Pulmonary embolism.
Conflict of interest statement
No potential conflicts of interest relevant to this article were reported.
Figures
Comment in
-
Comments on point of care D-dimer testing in the emergency department: a bioequivalence study.Ann Lab Med. 2014 Jan;34(1):64-5. doi: 10.3343/alm.2014.34.1.64. Epub 2013 Dec 6. Ann Lab Med. 2014. PMID: 24422199 Free PMC article. No abstract available.
-
Response to the comments on 'point of care D-dimer testing in the emergency department--a bioequivalence study' and erratum to the results.Ann Lab Med. 2014 Jan;34(1):66-7. doi: 10.3343/alm.2014.34.1.66. Epub 2013 Dec 6. Ann Lab Med. 2014. PMID: 24422200 Free PMC article. No abstract available.
References
-
- Lee-Lewandrowski E, Nichols J, Van Cott E, Grisson R, Louissaint A, Benzer T, et al. Implementation of a rapid whole blood D-dimer test in the emergency department of an urban academic medical center: impact on ED length of stay and ancillary test utilization. Am J Clin Pathol. 2009;132:326–331. - PubMed
-
- Walker JB, Nesheim ME. The molecular weights, mass distribution, chain composition, and structure of soluble fibrin degradation products released from a fibrin clot perfused with plasmin. J Biol Chem. 1999;274:5201–5212. - PubMed
-
- Greenberg CS, Devine DV, McCrae KM. Measurement of plasma fibrin D-dimer levels with the use of a monoclonal antibody coupled to latex beads. Am J Clin Pathol. 1987;87:94–100. - PubMed
-
- Dempfle CE. Validation, calibration, and specificity of quantitative D-dimer assays. Semin Vasc Med. 2005;5:315–320. - PubMed
-
- Pittet JL, de Moerloose P, Reber G, Durand C, Villard C, Piga N, et al. VIDAS D-dimer: fast quantitative ELISA for measuring D-dimer in plasma. Clin Chem. 1996;42:410–415. - PubMed
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous