Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Comparative Study
. 2013 Jul;22(7):940-7.
doi: 10.1016/j.jse.2012.09.007. Epub 2013 Jan 10.

Glenoid implant orientation and cement failure in total shoulder arthroplasty: a finite element analysis

Affiliations
Comparative Study

Glenoid implant orientation and cement failure in total shoulder arthroplasty: a finite element analysis

Charlie Yongpravat et al. J Shoulder Elbow Surg. 2013 Jul.

Abstract

Background: To minimize glenoid implant loosening in total shoulder arthroplasty (TSA), the ideal surgical procedure achieves correction to neutral version, complete implant-bone contact, and bone stock preservation. These goals, however, are not always achievable, and guidelines to prioritize their impact are not well established. The purpose of this study was to investigate how the degree of glenoid correction affects potential cement failure.

Methods: Eight patient-specific computer models were created for 4 TSA scenarios with different permutations of retroversion correction and implant-bone contact. Two bone models were used: a homogeneous cortical bone model and a heterogeneous cortical-trabecular bone model. A 750-N load was simulated, and cement stress was calculated. The risk of cement mantle fracture was reported as the percentage of cement stress exceeding the material endurance limit.

Results: Orienting the glenoid implant in retroversion resulted in the highest risk of cement fracture in a homogeneous bone model (P < .05). In the heterogeneous bone model, complete correction resulted in the highest risk of failure (P = .0028). A positive correlation (ρ = 0.901) was found between the risk of cement failure and amount of exposed trabecular bone.

Conclusions: Incorporating trabecular bone into the model changed the effect of implant orientation on cement failure. As exposed trabecular bone increased, the risk of cement fracture increased. This may be due to shifting the load-bearing support underneath the cement from cortical bone to trabecular bone.