Implant-bone load transfer mechanisms in complete-arch prostheses supported by four implants: a three-dimensional finite element approach
- PMID: 23328192
- DOI: 10.1016/S0022-3913(13)60004-9
Implant-bone load transfer mechanisms in complete-arch prostheses supported by four implants: a three-dimensional finite element approach
Abstract
Statement of problem: Complete-arch restorations supported by fewer than 5 dental implants can induce unbalanced load transfer and tissue overloading, leading to excessive bone resorption and possible clinical failure. This is primarily affected by the cantilever length, the implant design and positioning, and the morphology and properties of the bone.
Purpose: The purpose of this study was to compare 2 different restorative techniques for complete-arch rehabilitations supported by 4 implants. The primary purpose was to highlight the possible risks of excessive stress and unbalanced load transfer mechanisms and to identify the main biomechanical factors affecting loading transmission.
Material and methods: Three-dimensional (3D) numerical models of edentulous maxillae and mandibles restored with 2 techniques using 4 implants were generated from computed tomography (CT) images and analyzed with linear elastic finite-element simulations with 3 different static loads. The first technique used 2 vertical mesial implants and 2 tilted distal implants (at a 30 degree angle), and the second used vertical implants that fulfilled platform switching concepts. Bone-muscle interactions and temporomandibular joints were included in the mandibular model. Complete implant osseous integration was assumed and different posthealing crestal bone geometries were modeled. Stress measures (revealing risks of tissue overloading) and a performance index (highlighting the main features of the loading partition mechanisms) were introduced and computed to compare the 2 techniques.
Results: Dissimilar load transfer mechanisms of the 2 restorative approaches when applied in mandibular and maxillary models were modeled. Prostheses supported by distally tilted implants exhibited a more effective and uniform loading partition than all vertical implants, except in the simulated maxilla under a frontal load. Tilted distal implants reduced compressive states at distal bone-implant interfaces but, depending on bone morphology and loading type, could induce high tensile stresses at distal crests. Overloading risks on mesial periimplant bone decreased when the efficient preservation of the crestal bone through platform switching strategies was modeled.
Conclusions: Numerical simulations highlighted that the cantilever length, the implant design and positioning, and the bone's mechanical properties and morphology can affect both load transmission mechanisms and bone overloading risks in complete-arch restorations supported by 4 implants. Distally tilted implants induced better loading transmission than vertical implants, although the levels of computed stress were physiologically acceptable in both situations.
Copyright © 2013 The Editorial Council of the Journal of Prosthetic Dentistry. Published by Mosby, Inc. All rights reserved.
Similar articles
-
The influence of implant diameter and length on stress distribution of osseointegrated implants related to crestal bone geometry: a three-dimensional finite element analysis.J Prosthet Dent. 2008 Dec;100(6):422-31. doi: 10.1016/S0022-3913(08)60259-0. J Prosthet Dent. 2008. PMID: 19033026
-
Biomechanical comparison of axial and tilted implants for mandibular full-arch fixed prostheses.Int J Oral Maxillofac Implants. 2011 Sep-Oct;26(5):976-84. Int J Oral Maxillofac Implants. 2011. PMID: 22010079
-
Three-dimensional finite element analysis of vertical and angular misfit in implant-supported fixed prostheses.Int J Oral Maxillofac Implants. 2011 Jul-Aug;26(4):788-96. Int J Oral Maxillofac Implants. 2011. PMID: 21841989
-
Finite element stress analysis of cuneiform and cylindrical threaded implant geometries.Technol Health Care. 2006;14(4-5):421-38. Technol Health Care. 2006. PMID: 17065763 Review.
-
Stimulating effect of implant loading on surrounding bone. Comparison of three numerical models and validation by in vivo data.Clin Oral Implants Res. 2004 Apr;15(2):239-48. doi: 10.1111/j.1600-0501.2004.01000.x. Clin Oral Implants Res. 2004. PMID: 15008937 Review.
Cited by
-
Angulated implants: an alternative to bone augmentation and sinus lift procedure: systematic review.J Clin Diagn Res. 2015 Mar;9(3):ZE10-3. doi: 10.7860/JCDR/2015/11368.5655. Epub 2015 Mar 1. J Clin Diagn Res. 2015. PMID: 25954718 Free PMC article. Review.
-
Comparison between all-on-four and all-on-six treatment concepts on stress distribution for full-mouth rehabilitation using three-dimensional finite element analysis: A biomechanical study.J Indian Soc Periodontol. 2023 Mar-Apr;27(2):180-188. doi: 10.4103/jisp.jisp_278_22. Epub 2023 Mar 4. J Indian Soc Periodontol. 2023. PMID: 37152467 Free PMC article.
-
Finite element analysis of low-profile reconstruction plates for atrophic mandibles: a comparison of novel 3D grid and conventional plate designs.Oral Maxillofac Surg. 2024 Jun;28(2):595-603. doi: 10.1007/s10006-023-01173-3. Epub 2023 Jul 29. Oral Maxillofac Surg. 2024. PMID: 37516667
-
Evaluation of Biomechanical Effects of Mandible Arch Types in All-on-4 and All-on-5 Dental Implant Design: A 3D Finite Element Analysis.J Funct Biomater. 2025 Apr 7;16(4):134. doi: 10.3390/jfb16040134. J Funct Biomater. 2025. PMID: 40278242 Free PMC article.
-
Finite element analysis of the angle range in trans-inferior alveolar nerve implantation at the mandibular second molar.BMC Oral Health. 2023 Nov 25;23(1):928. doi: 10.1186/s12903-023-03641-4. BMC Oral Health. 2023. PMID: 38007495 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Research Materials