Real-time integration of MDCT-derived coronary anatomy and epicardial fat: impact on epicardial electroanatomic mapping and ablation for ventricular arrhythmias
- PMID: 23328560
- DOI: 10.1016/j.jcmg.2012.05.016
Real-time integration of MDCT-derived coronary anatomy and epicardial fat: impact on epicardial electroanatomic mapping and ablation for ventricular arrhythmias
Abstract
Objectives: This study aimed to evaluate the feasibility and accuracy of real-time integration of multidetector computed tomography (MDCT) derived coronary anatomy and epicardial fat distribution and its impact on electroanatomical mapping and ablation.
Background: Epicardial catheter ablation for ventricular arrhythmias (VA) is an important therapeutic option in patients after endocardial ablation failure. However, epicardial mapping and ablation are limited by the presence of coronary arteries and epicardial fat.
Methods: Twenty-eight patients (21 male, age 59 ± 16 years) underwent combined endo-epicardial electroanatomical mapping. Prior to the procedure, MDCT derived coronary anatomy and epicardial fat meshes were loaded into the mapping system (CARTO XP, Biosense Webster Inc, Diamond Bar, California). Real-time registration of MDCT data was performed after endocardial mapping. The distance between epicardial ablation sites and coronary arteries was assessed by registered MDCT and angiography. After the procedure, mapping and ablation points were superimposed on the MDCT using a reversed registration matrix for head-to-head comparison of mapping data and corresponding fat thickness.
Results: Image registration was successful and accurate in all patients (position error 2.8 ± 1.3 mm). At sites without evidence for scar, epicardial bipolar voltage decreased significantly (p < 0.001) with increasing fat thickness. Forty-six VA were targeted; 25 (54%) were abolished by catheter ablation, in 21 (46%) ablation failed. In 5 VA no target site was identified and in 3 VA adhesions prevented mapping. In 2 VA ablation was withheld due to His-bundle vicinity and in 7 VA due to proximity of coronary arteries. In 4 VA catheter ablation was ineffective. At ineffective ablation sites epicardial fat was significantly thicker compared to successful sites (16.9 ± 6.8 mm [range 7.3 to 22.2 mm] and 1.5 ± 2.1 mm [range 0.0 to 6.1 mm], p = 0.002).
Conclusions: Real-time image integration of pre-acquired MDCT information is feasible and accurate. Epicardial fat >7 mm and the presence of coronary arteries are important reasons for epicardial ablation failure. Visualization of fat thickness during the procedure may facilitate interpretation of bipolar electrograms and identification of ineffective ablation sites.
Copyright © 2013 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.
Comment in
-
The value of image integration for epicardial catheter ablation of ventricular tachycardia.JACC Cardiovasc Imaging. 2013 Jan;6(1):53-5. doi: 10.1016/j.jcmg.2012.06.009. JACC Cardiovasc Imaging. 2013. PMID: 23328561 No abstract available.
-
EP goes imaging.JACC Cardiovasc Imaging. 2013 Jan;6(1):133-5. doi: 10.1016/j.jcmg.2012.12.001. JACC Cardiovasc Imaging. 2013. PMID: 23328576 No abstract available.
Similar articles
-
Usefulness of epicardial impedance evaluation for epicardial mapping and determination of epicardial ablation site for ventricular tachycardia: A pilot study.J Cardiovasc Electrophysiol. 2018 Jan;29(1):138-145. doi: 10.1111/jce.13361. Epub 2017 Nov 1. J Cardiovasc Electrophysiol. 2018. PMID: 28988444
-
Image integration into 3-dimensional-electro-anatomical mapping system facilitates safe ablation of ventricular arrhythmias originating from the aortic root and its vicinity.Europace. 2018 Mar 1;20(3):520-527. doi: 10.1093/europace/euw399. Europace. 2018. PMID: 28340078
-
Integration of merged delayed-enhanced magnetic resonance imaging and multidetector computed tomography for the guidance of ventricular tachycardia ablation: a pilot study.J Cardiovasc Electrophysiol. 2013 Apr;24(4):419-26. doi: 10.1111/jce.12052. Epub 2012 Dec 17. J Cardiovasc Electrophysiol. 2013. PMID: 23252727
-
Epicardial Ablation via Arterial and Venous Systems.Card Electrophysiol Clin. 2020 Sep;12(3):391-399. doi: 10.1016/j.ccep.2020.06.006. Epub 2020 Jul 8. Card Electrophysiol Clin. 2020. PMID: 32771192 Review.
-
Epicardial approach for cardiac electrophysiology procedures.J Cardiovasc Electrophysiol. 2020 Jan;31(1):345-359. doi: 10.1111/jce.14282. Epub 2019 Dec 1. J Cardiovasc Electrophysiol. 2020. PMID: 31724787 Review.
Cited by
-
Unipolar voltage mapping in right ventricular cardiomyopathy: pitfalls, solutions and advantages.Europace. 2023 Mar 30;25(3):1035-1040. doi: 10.1093/europace/euac278. Europace. 2023. PMID: 36639881 Free PMC article.
-
Imaging-guided Ventricular Tachycardia Ablation.Arrhythm Electrophysiol Rev. 2013 Nov;2(2):128-34. doi: 10.15420/aer.2013.2.2.128. Epub 2013 Nov 29. Arrhythm Electrophysiol Rev. 2013. PMID: 26835054 Free PMC article.
-
The transmural activation interval: a new mapping tool to identify ventricular tachycardia substrates in right ventricular cardiomyopathy.Europace. 2023 Feb 16;25(2):478-486. doi: 10.1093/europace/euac220. Europace. 2023. PMID: 36480385 Free PMC article.
-
Catheter Ablation of Ventricular Tachycardia in Arrhythmogenic Right Ventricular Dysplasia/Cardiomyopathy.Korean Circ J. 2018 Oct;48(10):890-905. doi: 10.4070/kcj.2018.0268. Korean Circ J. 2018. PMID: 30238706 Free PMC article. Review.
-
Impact of electrode orientation, myocardial wall thickness, and myofiber direction on intracardiac electrograms: numerical modeling and analytical solutions.Front Physiol. 2023 Jul 10;14:1213218. doi: 10.3389/fphys.2023.1213218. eCollection 2023. Front Physiol. 2023. PMID: 37492643 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical