. 2013 Mar;39(3):343-9.
doi: 10.1016/j.jcrs.2012.09.017. Epub 2013 Jan 16.

Visual outcomes and subjective experience after bilateral implantation of a new diffractive trifocal intraocular lens

Affiliations

Visual outcomes and subjective experience after bilateral implantation of a new diffractive trifocal intraocular lens

Amy L Sheppard et al. J Cataract Refract Surg. 2013 Mar.

Abstract

Purpose: To assess clinical outcomes and subjective experience after bilateral implantation of a diffractive trifocal intraocular lens (IOL).

Setting: Midland Eye Institute, Solihull, United Kingdom.

Design: Cohort study.

Methods: Patients had bilateral implantation of Finevision trifocal IOLs. Uncorrected distance visual acuity, corrected distance visual acuity (CDVA), and manifest refraction were measured 2 months postoperatively. Defocus curves were assessed under photopic and mesopic conditions over a range of +1.50 to -4.00 diopters (D) in 0.50 D steps. Contrast sensitivity function was assessed under photopic conditions. Halometry was used to measure the angular size of monocular and binocular photopic scotomas arising from a glare source. Patient satisfaction with uncorrected near vision was assessed using the Near Activity Visual Questionnaire (NAVQ).

Results: The mean monocular CDVA was 0.08 logMAR ± 0.08 (SD) and the mean binocular CDVA, 0.06 ± 0.08 logMAR. Defocus curve testing showed an extended range of clear vision from +1.00 to -2.50 D defocus, with a significant difference in acuity between photopic conditions and mesopic conditions at -1.50 D defocus only. Photopic contrast sensitivity was significantly better binocularly than monocularly at all spatial frequencies. Halometry showed a glare scotoma of a mean size similar to that in previous studies of multifocal and accommodating IOLs; there were no subjective complaints of dysphotopsia. The mean NAVQ Rasch score for satisfaction with near vision was 15.9 ± 10.7 logits.

Conclusions: The trifocal IOL implanted binocularly produced good distance visual acuity and near and intermediate visual function. Patients were very satisfied with their uncorrected near vision.

PubMed Disclaimer

Similar articles

Cited by

Publication types