Randomized trial of pacemaker and lead system for safe scanning at 1.5 Tesla
- PMID: 23333721
- DOI: 10.1016/j.hrthm.2013.01.022
Randomized trial of pacemaker and lead system for safe scanning at 1.5 Tesla
Abstract
Background: Magnetic resonance imaging (MRI) of pacemakers is a relative contraindication because of the risks to the patient from potentially hazardous interactions between the MRI and the pacemaker system. Chest scans (ie, cardiac magnetic resonance scans) are of particular importance and higher risk. The previously Food and Drug Administration-approved magnetic resonance conditional system includes positioning restrictions, limiting the powerful utility of MRI.
Objective: To confirm the safety and effectiveness of a pacemaker system designed for safe whole body MRI without MRI scan positioning restrictions.
Methods: Primary eligibility criteria included standard dual-chamber pacing indications. Patients (n = 263) were randomized in a 2:1 ratio to undergo 16 chest and head scans at 1.5 T between 9 and 12 weeks postimplant (n = 177) or to not undergo MRI (n = 86) post-implant. Evaluation of the pacemaker system occurred immediately before, during (monitoring), and after MRI, 1-week post-MRI, and 1-month post-MRI, and similarly for controls. Primary end points measured the MRI-related complication-free rate for safety and compared pacing capture threshold between MRI and control subjects for effectiveness.
Results: There were no MRI-related complications during or after MRI in subjects undergoing MRI (n = 148). Differences in pacing capture threshold values from pre-MRI to 1-month post-MRI were minimal and similar between the MRI and control groups.
Conclusions: This randomized trial demonstrates that the Advisa MRI pulse generator and CapSureFix MRI 5086MRI lead system is safe and effective in the 1.5 T MRI environment without positioning restrictions for MRI scans or limitations of body parts scanned.
Copyright © 2013 Heart Rhythm Society. Published by Elsevier Inc. All rights reserved.
Similar articles
-
MRI scanning in patients with new and existing CapSureFix Novus 5076 pacemaker leads: randomized trial results.Heart Rhythm. 2015 Apr;12(4):759-65. doi: 10.1016/j.hrthm.2014.12.035. Epub 2014 Dec 30. Heart Rhythm. 2015. PMID: 25556038 Clinical Trial.
-
Magnetic resonance imaging in patients with a pacemaker system designed for the magnetic resonance environment.Heart Rhythm. 2011 Jan;8(1):65-73. doi: 10.1016/j.hrthm.2010.10.002. Epub 2010 Oct 7. Heart Rhythm. 2011. PMID: 20933098 Clinical Trial.
-
Real-World Evaluation of Magnetic Resonance Imaging in Patients With a Magnetic Resonance Imaging Conditional Pacemaker System: Results of 4-Year Prospective Follow-Up in 2,629 Patients.JACC Clin Electrophysiol. 2017 Nov;3(11):1231-1239. doi: 10.1016/j.jacep.2017.05.011. Epub 2017 Sep 13. JACC Clin Electrophysiol. 2017. PMID: 29759618
-
Magnetic resonance imaging in patients with cardiac pacemakers: era of "MR Conditional" designs.J Cardiovasc Magn Reson. 2011 Oct 27;13(1):63. doi: 10.1186/1532-429X-13-63. J Cardiovasc Magn Reson. 2011. PMID: 22032338 Free PMC article. Review.
-
Magnetic resonance imaging compatibility and safety of the SOUNDTEC Direct System.Laryngoscope. 2006 Aug;116(8):1321-33. doi: 10.1097/01.mlg.0000230479.39551.4a. Laryngoscope. 2006. PMID: 16885731 Review.
Cited by
-
Safety and performance of MR-conditional pacing systems with automated MRI mode at 1.5 and 3 Tesla.Eur Radiol. 2023 Oct;33(10):6948-6958. doi: 10.1007/s00330-023-09650-9. Epub 2023 May 17. Eur Radiol. 2023. PMID: 37195432 Free PMC article.
-
Safety of implanted cardiac devices in an MRI environment.Curr Cardiol Rep. 2015 Jul;17(7):605. doi: 10.1007/s11886-015-0605-9. Curr Cardiol Rep. 2015. PMID: 26026996 Review.
-
MRI-conditional pacemakers: current perspectives.Med Devices (Auckl). 2014 May 7;7:115-24. doi: 10.2147/MDER.S44063. eCollection 2014. Med Devices (Auckl). 2014. PMID: 24851058 Free PMC article. Review.
-
MRI for patients with cardiac implantable electrical devices.Cardiol Clin. 2014 May;32(2):299-304. doi: 10.1016/j.ccl.2013.12.002. Epub 2014 Feb 12. Cardiol Clin. 2014. PMID: 24793805 Free PMC article. Review.
-
What Is the Clinical Impact of Stress CMR After the ISCHEMIA Trial?Front Cardiovasc Med. 2021 Jun 4;8:683434. doi: 10.3389/fcvm.2021.683434. eCollection 2021. Front Cardiovasc Med. 2021. PMID: 34164444 Free PMC article. Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical