Radiation Exposure of the Hand and Chest during C-arm Fluoroscopy-Guided Procedures
- PMID: 23342208
- PMCID: PMC3546211
- DOI: 10.3344/kjp.2013.26.1.51
Radiation Exposure of the Hand and Chest during C-arm Fluoroscopy-Guided Procedures
Abstract
Background: The C-arm fluoroscope is an essential tool for the intervention of pain. The aim of this study was to investigate the radiation exposure experienced by the hand and chest of pain physicians during C-arm fluoroscopy-guided procedures.
Methods: This is a prospective study about radiation exposure to physicians during transforaminal epidural steroid injection (TFESI) and medial branch block (MBB). Four pain physicians were involved in this study. Data about effective dose (ED) at each physician's right hand and left side of the chest, exposure time, radiation absorbed dose (RAD), and the distance from the center of the X-ray field to the physician during X-ray scanning were collected.
Results: Three hundred and fifteen cases were included for this study. Demographic data showed no significant differences among the physicians in the TFESIs and MBBs. In the TFESI group, there was a significant difference between the ED at the hand and chest in all the physicians. In physician A, B and C, the ED at the chest was more than the ED at the hand. The distance from the center of the X-ray field to physician A was more than that of the other physicians, and for the exposure time, the ED and RAD in physician A was less than that of the other physicians. In the MBB group, there was no difference in the ED at the hand and chest, except for physician D. The distance from the center of the X-ray field to physician A was more than that of the other physicians and the exposure time in physician A was less than that of the other physicians.
Conclusions: In conclusion, the distance from the radiation source, position of the hand, experience and technique can correlate with the radiation dose.
Keywords: distance; exposure time; radiation dose; radiation protection.
Figures



Similar articles
-
The Radiation Exposure of Radiographer Related to the Location in C-arm Fluoroscopy-guided Pain Interventions.Korean J Pain. 2014 Apr;27(2):162-7. doi: 10.3344/kjp.2014.27.2.162. Epub 2014 Mar 28. Korean J Pain. 2014. PMID: 24748945 Free PMC article.
-
A Randomized Controlled Trial about the Levels of Radiation Exposure Depends on the Use of Collimation C-arm Fluoroscopic-guided Medial Branch Block.Korean J Pain. 2013 Apr;26(2):148-53. doi: 10.3344/kjp.2013.26.2.148. Epub 2013 Apr 3. Korean J Pain. 2013. PMID: 23614076 Free PMC article.
-
Does nonexistent of your hands on the screen guarantee no radiation exposure to your body? - Study on exposure of the practitioner's hands to radiation during C-arm fluoroscopy-guided injections and effectiveness of a new shielding device.Medicine (Baltimore). 2019 Nov;98(46):e17959. doi: 10.1097/MD.0000000000017959. Medicine (Baltimore). 2019. PMID: 31725656 Free PMC article.
-
Radiation Exposure in Interventional Pain Management Physicians: A Systematic Review of the Current Literature.Pain Physician. 2024 Jan;27(1):E17-E35. Pain Physician. 2024. PMID: 38285025
-
Radiation safety for pain physicians: principles and recommendations.Korean J Pain. 2022 Apr 1;35(2):129-139. doi: 10.3344/kjp.2022.35.2.129. Korean J Pain. 2022. PMID: 35354676 Free PMC article. Review.
Cited by
-
Radiation exposure to the eyes and thyroid during C-arm fluoroscopy-guided cervical epidural injections is far below the safety limit.Korean J Pain. 2020 Jan 1;33(1):73-80. doi: 10.3344/kjp.2020.33.1.73. Korean J Pain. 2020. PMID: 31888321 Free PMC article.
-
Radiation Protection in Interventional Radiology.Indian J Radiol Imaging. 2022 Jan 10;31(4):939-945. doi: 10.1055/s-0041-1741049. eCollection 2021 Oct. Indian J Radiol Imaging. 2022. PMID: 35136507 Free PMC article. Review.
-
Are Doctors Exposed to Radiation Even When Wearing Protectors during Fluoroscopic Procedures?Korean J Pain. 2013 Apr;26(2):207. doi: 10.3344/kjp.2013.26.2.207. Epub 2013 Apr 3. Korean J Pain. 2013. PMID: 23614089 Free PMC article. No abstract available.
-
Comparison of Efficacy between Transforaminal Epidural Steroid Injection Technique without Contrast versus with Contrast in Lumbar Radiculopathy: A Prospective Longitudinal Cohort Study.Asian Spine J. 2023 Dec;17(6):1108-1116. doi: 10.31616/asj.2023.0007. Epub 2023 Nov 10. Asian Spine J. 2023. PMID: 37946336 Free PMC article.
-
The Radiation Exposure of Radiographer Related to the Location in C-arm Fluoroscopy-guided Pain Interventions.Korean J Pain. 2014 Apr;27(2):162-7. doi: 10.3344/kjp.2014.27.2.162. Epub 2014 Mar 28. Korean J Pain. 2014. PMID: 24748945 Free PMC article.
References
-
- Miller DL, Vañó E, Bartal G, Balter S, Dixon R, Padovani R, et al. Occupational radiation protection in interventional radiology: a joint guideline of the Cardiovascular and Interventional Radiology Society of Europe and the Society of Interventional Radiology. J Vasc Interv Radiol. 2010;21:607–615. - PubMed
-
- Harding LK, Thomson WH. International commission on radiation protection. Nucl Med Commun. 1990;11:585–587. - PubMed
LinkOut - more resources
Full Text Sources
Other Literature Sources