Equally good fixation of cemented and uncemented cups in total trapeziometacarpal joint prostheses. A randomized clinical RSA study with 2-year follow-up
- PMID: 23343372
- PMCID: PMC3584612
- DOI: 10.3109/17453674.2013.765625
Equally good fixation of cemented and uncemented cups in total trapeziometacarpal joint prostheses. A randomized clinical RSA study with 2-year follow-up
Abstract
Background and purpose: Cup failure is a recognized problem in total trapeziometacarpal (TM) joint prostheses; it may be related to poor fixation, which can be revealed by radiostereometric analysis (RSA). We compared the early implant migration of cemented trapezium cups to that of uncemented screw cups.
Patients and methods: In a prospective, parallel-group, randomized patient-blinded clinical trial, we included 32 hands in 28 patients (5 males) with a mean age of 58 (40-77) years and with Eaton stage-2 or -3 osteoarthritis of the trapeziometacarpal joint. Patients were randomized to surgery with a cemented DLC all-polyethylene cup (C) (n = 16) or an uncemented hydroxyapatite-coated chrome-cobalt Elektra screw cup (UC) (n = 16). Uncemented cups were inserted without threading of the bone. Stereoradiographs for evaluation of cup migration (primary effect size) and DASH and pain scores were obtained during 2 years of follow-up.
Results: The 2-year total translation (TT) was similar (p = 0.2): 0.24 mm (SD 0.10) for the C (n = 11) and 0.19 mm (SD 0.16) for the UC (n = 11). Variances were similar (p = 0.4). Judged by RSA, 2 UC cups and 1 C cup became loose (TT > 1 mm). Both UC cups were found to be loose at revision. Grip strength, pain, and DASH scores were similar between groups at all measurement points.
Interpretation: Early implant fixation and clinical outcome were equally good with both cup designs. This is the first clinical RSA study on trapezium cups, and the method appears to be clinically useful for detection of loose implants.
Figures
References
-
- Aspenberg P, Wagner P, Nilsson KG, Ranstam J. Fixed or loose? Dochotomy in RSA data for cemented cups. Acta Orthop. 2008;79(4):467–73. - PubMed
-
- Braun RM. Total joint replacement at the base of the thumb – preliminary report. J Hand Surg Am. 1982;7:245–51. - PubMed
-
- Bryant D, Havey TC, Roberts R, Guyatt G. How many patients? How many limbs? Analysis of patients or limbs in the orthopaedic literature: A systematic review. J Bone Joint Surg (Am) 2006;88(1):41–5. - PubMed
-
- Cooney WP, Chao EY. Biomechanical analysis of static forces in the thumb during hand function. J Bone Joint Surg (Am) 1977;59:27–36. - PubMed
-
- Cooney WP, Linscheid RP, Askew LJ. Total arthroplasty of the thumb trapeziometacarpal joint. Clin Orthop. 1987;220:35–45. - PubMed
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources