Effect of x-ray tube parameters and iodine concentration on image quality and radiation dose in cerebral pediatric and adult CT angiography: a phantom study
- PMID: 23344518
- DOI: 10.1097/RLI.0b013e31827efc17
Effect of x-ray tube parameters and iodine concentration on image quality and radiation dose in cerebral pediatric and adult CT angiography: a phantom study
Abstract
Objectives: The aim of the present phantom study was to investigate the effect of x-ray tube parameters and iodine concentration on image quality and radiation dose in cerebral computed tomographic (CT) angiographic examinations of pediatric and adult individuals.
Materials and methods: Four physical anthropomorphic phantoms that represent the average individual as neonate, 1-year-old, 5-year-old, and 10-year-old children and the RANDO phantom that simulates the average adult individual were used. Cylindrical vessels were bored along the brain-equivalent plugs of each physical phantom. To simulate the brain vasculature, vessels of 0.6, 1, 2, and 3 mm in diameter were created. These vessels were filled with contrast medium (CM) solutions at different iodine concentrations, that is, 5.6, 4.2, 2.7, and 1.4 mg I/mL. The phantom heads were scanned at 120, 100, and 80 kV. The applied quality reference tube current-time product values ranged from a minimum of 45 to a maximum of 680. The CT acquisitions were performed on a 16-slice CT scanner using the automatic exposure control system. Image quality was evaluated on the basis of image noise and contrast-to-noise ratio (CNR) between the contrast-enhanced iodinated vessels and the unenhanced regions of interest. Dose reduction was calculated as the percentage difference of the CT dose index value at the quality reference tube current-time product and the CT dose index at the mean modulated tube current-time product.
Results: Image noise that was measured using the preset tube current-time product settings varied significantly among the different phantoms (P < 0.0001). Hounsfield unit number of iodinated vessels was linearly related to CM concentration (r² = 0.907) and vessel diameter (r² = 0.918). The Hounsfield unit number of iodinated vessels followed a decreasing trend from the neonate phantom to the adult phantom at all kilovoltage settings. For the same image noise level, a CNR improvement of up to 69% and a dose reduction of up to 61% may be achieved when CT acquisition is performed at 80 kV compared with 120 kV. For the same CNR, a reduction by 25% of the administered CM concentration may be achieved when CT acquisition is performed at 80 kV compared with 120 kV.
Conclusions: In cerebral CT angiographic studies, appropriate adjustment of the preset tube current-time product settings is required to achieve the same image noise level among participants of different age. Cerebral CT angiography at 80 kV significantly improves CNR and significantly reduces radiation dose. Moreover, at 80 kV, a considerable reduction of the administered amount of the CM may be reached, thus reducing potential risks for contrast-induced nephropathy.
Similar articles
-
Effect of X-ray tube parameters, iodine concentration, and patient size on image quality in pulmonary computed tomography angiography: a chest-phantom-study.Invest Radiol. 2008 Jun;43(6):374-81. doi: 10.1097/RLI.0b013e3181690042. Invest Radiol. 2008. PMID: 18496042
-
Automated low-kilovoltage selection in pediatric computed tomography angiography: phantom study evaluating effects on radiation dose and image quality.Invest Radiol. 2013 Aug;48(8):584-9. doi: 10.1097/RLI.0b013e318289f918. Invest Radiol. 2013. PMID: 23563195
-
Pediatric thoracic CT angiography at 70 kV: a phantom study to investigate the effects on image quality and radiation dose.Pediatr Radiol. 2016 Jul;46(8):1114-9. doi: 10.1007/s00247-016-3588-9. Epub 2016 Mar 17. Pediatr Radiol. 2016. PMID: 26987734
-
Optimal tube potential for radiation dose reduction in pediatric CT: principles, clinical implementations, and pitfalls.Radiographics. 2011 May-Jun;31(3):835-48. doi: 10.1148/rg.313105079. Radiographics. 2011. PMID: 21571660 Review.
-
Low radiation dose in computed tomography: the role of iodine.Br J Radiol. 2017 Aug;90(1076):20170079. doi: 10.1259/bjr.20170079. Epub 2017 May 4. Br J Radiol. 2017. PMID: 28471242 Free PMC article. Review.
Cited by
-
Automatic Tube Current Modulation and Tube Voltage Selection in Pediatric Computed Tomography: A Phantom Study on Radiation Dose and Image Quality.Invest Radiol. 2019 May;54(5):265-272. doi: 10.1097/RLI.0000000000000537. Invest Radiol. 2019. PMID: 30562273 Free PMC article.
-
A comprehensive assessment of physical image quality of five different scanners for head CT imaging as clinically used at a single hospital centre-A phantom study.PLoS One. 2021 Jan 14;16(1):e0245374. doi: 10.1371/journal.pone.0245374. eCollection 2021. PLoS One. 2021. PMID: 33444367 Free PMC article.
-
Radiation doses with various body weights of phantoms in brain 128-slice MDCT examination.J Radiat Res. 2019 Jul 1;60(4):466-475. doi: 10.1093/jrr/rrz029. J Radiat Res. 2019. PMID: 31197311 Free PMC article.
-
The Impact of Combining a Low-Tube Voltage Acquisition with Iterative Reconstruction on Total Iodine Dose in Coronary CT Angiography.Biomed Res Int. 2017;2017:2476171. doi: 10.1155/2017/2476171. Epub 2017 May 23. Biomed Res Int. 2017. PMID: 28620616 Free PMC article.
-
Contrast agent and radiation dose reduction in abdominal CT by a combination of low tube voltage and advanced image reconstruction algorithms.Eur Radiol. 2015 Apr;25(4):1023-31. doi: 10.1007/s00330-014-3510-5. Epub 2014 Nov 29. Eur Radiol. 2015. PMID: 25432293 Free PMC article.