Intra and inter-user reliability of central corneal thickness measurements obtained in healthy feline eyes using a portable spectral-domain optical coherence tomography device
- PMID: 23356722
- DOI: 10.1111/vop.12029
Intra and inter-user reliability of central corneal thickness measurements obtained in healthy feline eyes using a portable spectral-domain optical coherence tomography device
Abstract
Objective: To determine the intra and inter-user reliability of central corneal thickness (CCT) measurements obtained using a portable spectral-domain optical coherence tomography (SD-OCT) device in healthy feline eyes.
Animals: Twenty clinically normal cats.
Procedures: Gentle manual restraint was utilized for proper animal positioning. Central corneal thickness measurements were obtained from both eyes of each animal three times by two operators in succession. A 6 mm, corneal pachymetry protocol was performed using the Optovue® iVue SD-OCT system. The CCT was automatically calculated using the pachymetry software generated from eight radial scans, 6 mm in length. The CCT was recorded for each operator and compared statistically for both intra and inter-user reliability.
Results: Twenty cats (11 CM, 9 SF) with a mean age of 4.7 years were examined. The mean CCT for all eyes examined was 629.08 ± 47.05 μm. No significant difference in CCT was found based on the eye examined (OD vs. OS) or age of the animal. However, the CCT of spayed females was significantly higher (42.64 ± 18.0 μm) as compared to castrated males (P = 0.03). There was no significant difference in CCT measurements between replicates performed by the same operator or between operators. The mean difference in CCT between operators was 0.38 μm. The coefficient of variation for each user and between users was very low (range 0.68-1.5%). The intra-class correlation coefficient comparing operators was 0.975.
Conclusions: The Optovue® iVue SD-OCT device is capable of obtaining CCT measurements with excellent intra and interoperator reliability in healthy feline eyes.
Keywords: central corneal thickness; feline; optical coherence tomography; pachymetry.
© 2013 American College of Veterinary Ophthalmologists.
Similar articles
-
A spectral-domain optical coherence tomography device provides reliable corneal pachymetry measurements in canine eyes.Vet Rec. 2013 Jun 8;172(23):605. doi: 10.1136/vr.101530. Epub 2013 May 28. Vet Rec. 2013. PMID: 23716535
-
Manual corneal thickness measurements of healthy equine eyes using a portable spectral-domain optical coherence tomography device.Equine Vet J. 2014 Sep;46(5):631-4. doi: 10.1111/evj.12198. Epub 2013 Dec 9. Equine Vet J. 2014. PMID: 24131285
-
Central corneal thickness measurements in normal dogs: a comparison between ultrasound pachymetry and optical coherence tomography.Vet Ophthalmol. 2014 May;17(3):207-11. doi: 10.1111/vop.12074. Epub 2013 Jun 14. Vet Ophthalmol. 2014. PMID: 23763504
-
Reliability of manual measurements of corneal thickness obtained from healthy canine eyes using spectral-domain optical coherence tomography (SD-OCT).Can J Vet Res. 2014 Jul;78(3):221-5. Can J Vet Res. 2014. PMID: 24982554 Free PMC article.
-
Comparison between Pentacam-HR and optical coherence tomographycentral corneal thickness measurements in healthy feline eyes.Vet Ophthalmol. 2016 Jul;19 Suppl 1:105-14. doi: 10.1111/vop.12397. Vet Ophthalmol. 2016. PMID: 27370363
Cited by
-
Anterior and posterior segment parameters measured with Fourier domain optical coherence tomography in photopic and scotopic conditions.Indian J Ophthalmol. 2016 Feb;64(2):136-9. doi: 10.4103/0301-4738.179726. Indian J Ophthalmol. 2016. PMID: 27050349 Free PMC article.
-
Intraoperative pachymetry using spectral-domain optical coherence tomography during accelerated corneal collagen crosslinking.Biomed Res Int. 2013;2013:848363. doi: 10.1155/2013/848363. Epub 2013 Jul 25. Biomed Res Int. 2013. PMID: 23984411 Free PMC article. Clinical Trial.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous