Abstract
Background: Noninvasive prenatal detection of common fetal aneuploidies with cell-free DNA from maternal plasma has been achieved with high-throughput next-generation sequencing platforms. Turnaround times for previously tested platforms are still unsatisfactory for clinical applications, however, because of the time spent on sequencing. The development of semiconductor sequencing technology has provided a way to shorten overall run times. We studied the feasibility of using semiconductor sequencing technology for the noninvasive detection of fetal aneuploidy.
Methods: Maternal plasma DNA from 13 pregnant women, corresponding to 4 euploid, 6 trisomy 21 (T21), 2 trisomy 18 (T18), and 1 trisomy 13 (T13) pregnancies, were sequenced on the Ion Torrent Personal Genome Machine sequencer platform with 318 chips. The data were analyzed with the T statistic method after correcting for GC bias, and the T value was calculated as an indicator of fetal aneuploidy.
Results: We obtained a mean of 3 524 401 high-quality reads per sample, with an efficiency rate of 77.9%. All of the T21, T13, and T18 fetuses could be clearly distinguished from euploid fetuses, and the time spent on library preparation and sequencing was 24 h.
Conclusions: Semiconductor sequencing represents a suitable technology for the noninvasive prenatal detection of fetal aneuploidy. With this platform, sequencing times can be substantially reduced; however, a further larger-scale study is needed to determine the imprecision of noninvasive fetal aneuploidy detection with this system.
Comment in
-
Rapid prenatal testing using semiconductor sequencing?Clin Chem. 2013 May;59(5):735-6. doi: 10.1373/clinchem.2013.203299. Epub 2013 Mar 7. Clin Chem. 2013. PMID: 23471098 No abstract available.
Similar articles
-
Noninvasive prenatal screening for fetal trisomies 21, 18, 13 and the common sex chromosome aneuploidies from maternal blood using massively parallel genomic sequencing of DNA.Am J Obstet Gynecol. 2014 Oct;211(4):365.e1-12. doi: 10.1016/j.ajog.2014.03.042. Epub 2014 Mar 19. Am J Obstet Gynecol. 2014. PMID: 24657131
-
Open source non-invasive prenatal testing platform and its performance in a public health laboratory.Prenat Diagn. 2016 Jun;36(6):530-6. doi: 10.1002/pd.4819. Epub 2016 Apr 24. Prenat Diagn. 2016. PMID: 27027563
-
The feasibility study of non-invasive fetal trisomy 18 and 21 detection with semiconductor sequencing platform.PLoS One. 2014 Oct 20;9(10):e110240. doi: 10.1371/journal.pone.0110240. eCollection 2014. PLoS One. 2014. PMID: 25329639 Free PMC article.
-
Non-invasive prenatal testing for fetal chromosomal abnormalities by low-coverage whole-genome sequencing of maternal plasma DNA: review of 1982 consecutive cases in a single center.Ultrasound Obstet Gynecol. 2014 Mar;43(3):254-64. doi: 10.1002/uog.13277. Epub 2014 Feb 10. Ultrasound Obstet Gynecol. 2014. PMID: 24339153 Review.
-
Analysis of cell-free DNA in maternal blood in screening for fetal aneuploidies: updated meta-analysis.Ultrasound Obstet Gynecol. 2015 Mar;45(3):249-66. doi: 10.1002/uog.14791. Epub 2015 Feb 1. Ultrasound Obstet Gynecol. 2015. Update in: Ultrasound Obstet Gynecol. 2017 Sep;50(3):302-314. doi: 10.1002/uog.17484. PMID: 25639627 Updated. Review.
Cited by
-
MaterniCode: New Bioinformatic Pipeline to Detect Fetal Aneuploidies and Rearrangements Using Next-Generation Sequencing.Int J Genomics. 2024 Jun 13;2024:8859058. doi: 10.1155/2024/8859058. eCollection 2024. Int J Genomics. 2024. PMID: 38962150 Free PMC article.
-
Noninvasive Prenatal Testing: Comparison of Two Mappers and Influence in the Diagnostic Yield.Biomed Res Int. 2018 Jun 7;2018:9498140. doi: 10.1155/2018/9498140. eCollection 2018. Biomed Res Int. 2018. PMID: 29977923 Free PMC article.
-
Genomics-based non-invasive prenatal testing for detection of fetal chromosomal aneuploidy in pregnant women.Cochrane Database Syst Rev. 2017 Nov 10;11(11):CD011767. doi: 10.1002/14651858.CD011767.pub2. Cochrane Database Syst Rev. 2017. PMID: 29125628 Free PMC article.
-
Utilization of Benchtop Next Generation Sequencing Platforms Ion Torrent PGM and MiSeq in Noninvasive Prenatal Testing for Chromosome 21 Trisomy and Testing of Impact of In Silico and Physical Size Selection on Its Analytical Performance.PLoS One. 2015 Dec 15;10(12):e0144811. doi: 10.1371/journal.pone.0144811. eCollection 2015. PLoS One. 2015. PMID: 26669558 Free PMC article.
-
Comparison of two high-throughput semiconductor chip sequencing platforms in noninvasive prenatal testing for Down syndrome in early pregnancy.BMC Med Genomics. 2016 Apr 30;9(1):22. doi: 10.1186/s12920-016-0182-9. BMC Med Genomics. 2016. PMID: 27129388 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous