Real-time cardiorespiratory coherence detects antinociception during general anesthesia
- PMID: 23366759
- DOI: 10.1109/EMBC.2012.6346798
Real-time cardiorespiratory coherence detects antinociception during general anesthesia
Abstract
Heart rate variability (HRV) may provide anesthesiologists with a noninvasive tool for monitoring nociception during general anesthesia. A novel real-time cardiorespiratory coherence (CRC) algorithm has been developed to analyze the strength of linear coupling between heart rate (HR) and respiration. CRC values range from 0 (low coherence, strong nociception) to 1 (high coherence, no nociception). The algorithm uses specially designed filters to operate in real-time, minimizing computational complexity and time delay. In the standard HRV high frequency band of 0.15 - 0.4 Hz, the real-time delay is only 5.25 - 3.25 s. We have assessed the algorithm's response to 60 anesthetic bolus events (a large dose of anesthetics given over a short time; strongly antinociceptive) recorded in 47 pediatric patients receiving general anesthesia. Real-time CRC responded strongly to bolus events, changing by an average of 30%. For comparison, three traditional measures of HRV (LF/HF ratio, SDNN, and RMSSD) responded on average by only 3.8%, 14%, and 3.9%, respectively. Finally, two traditional clinical measures of nociception (HR and blood pressure) responded on average by only 3.9% and 0.91%, respectively. CRC may thus be used as a real-time nociception monitor during general anesthesia.