. 2013 Jun;48(6):359-65.
doi: 10.1097/RLI.0b013e3182806904.

Magnetic resonance imaging-guided focused ultrasound treatment of symptomatic uterine fibroids: impact of technology advancement on ablation volumes in 115 patients

Affiliations

Magnetic resonance imaging-guided focused ultrasound treatment of symptomatic uterine fibroids: impact of technology advancement on ablation volumes in 115 patients

Christoph G Trumm et al. Invest Radiol. 2013 Jun.

Abstract

Objectives: The aim of this study was to assess the impact of the advanced technology of the new ExAblate 2100 system (Insightec Ltd, Haifa, Israel) for magnetic resonance imaging (MRI)-guided focused ultrasound surgery on treatment outcomes in patients with symptomatic uterine fibroids, as measured by the nonperfused volume ratio.

Materials and methods: This is a retrospective analysis of 115 women (mean age, 42 years; range, 27-54 years) with symptomatic fibroids who consecutively underwent MRI-guided focused ultrasound treatment in a single center with the new generation ExAblate 2100 system from November 2010 to June 2011. Mean ± SD total volume and number of treated fibroids (per patient) were 89 ± 94 cm and 2.2 ± 1.7, respectively. Patient baseline characteristics were analyzed regarding their impact on the resulting nonperfused volume ratio.

Results: Magnetic resonance imaging-guided focused ultrasound treatment was technically successful in 115 of 123 patients (93.5%). In 8 patients, treatment was not possible because of bowel loops in the beam pathway that could not be mitigated (n = 6), patient movement (n = 1), and system malfunction (n = 1). Mean nonperfused volume ratio was 88% ± 15% (range, 38%-100%). Mean applied energy level was 5400 ± 1200 J, and mean number of sonications was 74 ± 27. No major complications occurred. Two cases of first-degree skin burn resolved within 1 week after the intervention. Of the baseline characteristics analyzed, only the planned treatment volume had a statistically significant impact on nonperfused volume ratio.

Conclusions: With technological advancement, the outcome of MRI-guided focused ultrasound treatment in terms of the nonperfused volume ratio can be enhanced with a high safety profile, markedly exceeding results reported in previous clinical trials.

PubMed Disclaimer

MeSH terms

LinkOut - more resources