. 2013 Sep;17(3):1109-15.
doi: 10.1007/s11325-013-0812-9. Epub 2013 Feb 6.

Sleep disordered breathing analysis in a general population using standard pulse oximeter signals

Affiliations

Sleep disordered breathing analysis in a general population using standard pulse oximeter signals

Deganit Barak-Shinar et al. Sleep Breath. 2013 Sep.

Abstract

Objectives: Obstructive sleep apnea reported as the apnea-hypopnea index (AHI) is usually measured in sleep laboratories using a high number of electrodes connected to the patient's body. In this study, we examined the use of a standard pulse oximeter system with an automated analysis based on the photoplethysmograph (PPG) signal for the diagnosis of sleep disordered breathing. Using a standard and simple device with high accuracy might provide a convenient diagnostic or screening solution for patient evaluation at home or in other out of center testing environments.

Methods: The study included 140 consecutive patients that were referred routinely to a sleep laboratory [SleepMed Inc.] for the diagnosis of sleep disordered breathing. Each patient underwent an overnight polysomnography (PSG) study according to AASM guidelines in an AASM-accredited sleep laboratory. The automatic analysis is based on photoplethysmographic and saturation signals only. Those two signals were recorded for the entire night as part of the full overnight PSG sleep study. The AHI calculated from the PPG analysis is compared to the AHI calculated from the manual scoring gold standard full PSG.

Results: The AHI and total respiratory events measured by the pulse oximeter analysis correlated very well with the corresponding results obtained by the gold standard full PSG. The sensitivity and specificity of AHI = or > 5 and 15 levels measured by the analysis are both above 90 %. The sensitivity and positive predictive value for the detection of respiratory event are both above 84 %.

Conclusions: The tested system in this study yielded an acceptable result of sleep disordered breathing compared to the gold standard PSG in patients with moderate to severe sleep apnea. Accordingly and given the convenience and simplicity of the standard pulse oximeter device, the new system can be considered suitable for home and ambulatory diagnosis or screening of sleep disordered breathing patients.

PubMed Disclaimer

Similar articles

Cited by

References

    1. J Clin Sleep Med. 2008 Feb 15;4(1):26-37 - PubMed
    1. Am J Respir Crit Care Med. 2001 Jan;163(1):19-25 - PubMed
    1. Chest. 1992 May;101(5):1221-7 - PubMed
    1. J Clin Nurs. 2007 Feb;16(2):336-43 - PubMed
    1. Sleep Breath. 2013 May;17(2):679-86 - PubMed

Publication types

LinkOut - more resources