Evolution of the ablation region after magnetic resonance-guided high-intensity focused ultrasound ablation in a Vx2 tumor model
- PMID: 23399810
- DOI: 10.1097/RLI.0b013e3182820257
Evolution of the ablation region after magnetic resonance-guided high-intensity focused ultrasound ablation in a Vx2 tumor model
Abstract
Objectives: Volumetric magnetic resonance (MR)-guided high-intensity focused ultrasound (HIFU) is a completely noninvasive image-guided thermal ablation technique. Recently, there has been growing interest in the use of MR-HIFU for noninvasive ablation of malignant tumors. Of particular interest for noninvasive ablation of malignant tumors is reliable treatment monitoring and evaluation of response. At this point, there is limited evidence on the evolution of the ablation region after MR-HIFU treatment. The purpose of the present study was to comprehensively characterize the evolution of the ablation region after volumetric MR-HIFU ablation in a Vx2 tumor model using MR imaging, MR temperature data, and histological data.
Materials and methods: Vx2 tumors in the hind limb muscle of New Zealand White rabbits (n = 30) were ablated using a clinical MR-HIFU system. Twenty-four animals were available for analyses. Magnetic resonance imaging was performed before and immediately after ablation; MR temperature mapping was performed during the ablation. The animals were distributed over 7 groups with different follow-up lengths. Depending on the group, animals were reimaged and then killed on day 0, 1, 3, 7, 14, 21, or 28 after ablation. For all time points, the size of nonperfused areas (NPAs) on contrast-enhanced T1-weighted (CE-T1-w) images was compared with lethal thermal dose areas (ie, the tissue area that received a thermal dose of 240 equivalent minutes or greater [EM] at 43°C) and with the necrotic tissue areas on histology sections.
Results: The NPA on CE-T1-w imaging showed an increase in median size from 266 ± 148 to 392 ± 178 mm(2) during the first day and to 343 ± 170 mm(2) on day 3, followed by a gradual decrease to 113 ± 103 mm(2) on day 28. Immediately after ablation, the NPA was 1.6 ± 1.4 times larger than the area that received a thermal dose of 240 EM or greater in all animals. The median size of the necrotic area on histology was 1.7 ± 0.4 times larger than the NPA immediately after ablation. After 7 days, the size of the NPA was in agreement with the necrotic tissue area on histology (ratio, 1.0 ± 0.2).
Conclusions: During the first 3 days after MR-HIFU ablation, the ablation region increases in size, after which it gradually decreases in size. The NPA on CE-T1-w imaging underestimates the extent of tissue necrosis on histology in the initial few days, but after 1 week, the NPA is reliable in delineating the necrotic tissue area. The 240-EM thermal dose limit underestimates the necrotic tissue area immediately after MR-HIFU ablation. Reliable treatment evaluation techniques are particularly important for noninvasive, image-guided tumor ablation. Our results indicate that CE-T1-w imaging is reliable for MR-HIFU treatment evaluation after 1 week.
Similar articles
-
Stability and trapping of magnetic resonance imaging contrast agents during high-intensity focused ultrasound ablation therapy.Invest Radiol. 2013 Jul;48(7):517-24. doi: 10.1097/RLI.0b013e31829aae98. Invest Radiol. 2013. PMID: 23695082
-
A clinically feasible treatment protocol for magnetic resonance-guided high-intensity focused ultrasound ablation in the liver.Invest Radiol. 2015 Jan;50(1):24-31. doi: 10.1097/RLI.0000000000000091. Invest Radiol. 2015. PMID: 25198833
-
The effects of Magnetic Resonance Imaging-guided High-Intensity Focused Ultrasound ablation on human cadaver breast tissue.Eur J Pharmacol. 2013 Oct 5;717(1-3):21-30. doi: 10.1016/j.ejphar.2012.11.070. Epub 2013 Apr 9. Eur J Pharmacol. 2013. PMID: 23583321
-
MRI methods for the evaluation of high intensity focused ultrasound tumor treatment: Current status and future needs.Magn Reson Med. 2016 Jan;75(1):302-17. doi: 10.1002/mrm.25758. Epub 2015 Jun 22. Magn Reson Med. 2016. PMID: 26096859 Review.
-
MR-guided high-intensity focused ultrasound ablation of breast cancer with a dedicated breast platform.Cardiovasc Intervent Radiol. 2013 Apr;36(2):292-301. doi: 10.1007/s00270-012-0526-6. Epub 2012 Dec 12. Cardiovasc Intervent Radiol. 2013. PMID: 23232856 Review.
Cited by
-
Multiparametric MRI analysis for the identification of high intensity focused ultrasound-treated tumor tissue.PLoS One. 2014 Jun 13;9(6):e99936. doi: 10.1371/journal.pone.0099936. eCollection 2014. PLoS One. 2014. PMID: 24927280 Free PMC article.
-
Procedural sedation and analgesia for respiratory-gated MR-HIFU in the liver: a feasibility study.J Ther Ultrasound. 2016 Jul 29;4:19. doi: 10.1186/s40349-016-0063-x. eCollection 2016. J Ther Ultrasound. 2016. PMID: 27478615 Free PMC article.
-
Application of Volumetric Modulated Arc Therapy and Simultaneous Integrated Boost Techniques to Prepare "Safe Margin" in the Rabbit VX2 Limb Tumor Model.Med Sci Monit. 2015 Aug 15;21:2397-405. doi: 10.12659/MSM.894909. Med Sci Monit. 2015. PMID: 26280694 Free PMC article.
-
CT spectral imaging for monitoring the therapeutic efficacy of VEGF receptor kinase inhibitor AG-013736 in rabbit VX2 liver tumours.Eur Radiol. 2017 Mar;27(3):918-926. doi: 10.1007/s00330-016-4458-4. Epub 2016 Jun 10. Eur Radiol. 2017. PMID: 27287476
-
Development of a high-field MR-guided HIFU setup for thermal and mechanical ablation methods in small animals.J Ther Ultrasound. 2015 Aug 13;3:14. doi: 10.1186/s40349-015-0035-6. eCollection 2015. J Ther Ultrasound. 2015. PMID: 26269744 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources