Optimizing positron emission tomography image acquisition protocols in integrated positron emission tomography/magnetic resonance imaging
- PMID: 23399811
- DOI: 10.1097/RLI.0b013e3182823695
Optimizing positron emission tomography image acquisition protocols in integrated positron emission tomography/magnetic resonance imaging
Abstract
Objectives: In integrated positron emission tomography (PET)/magnetic resonance imaging (MRI), the PET data acquisition is performed simultaneously to the magnetic resonance data acquisition, leaving latitude for the duration of PET acquisition time. This establishes emission time as an important parameter in forthcoming PET/MRI protocols because it is one of the key factors determining PET image quality. Thus, the purpose of the current study was to identify optimal duration of PET acquisition time in PET/MRI.
Materials and methods: A total of 22 consecutive patients (7 men, 15 women) underwent fluorine-18-labeled fluorodeoxyglucose (18[F]FDG) PET/MRI after clinical PET/computed tomography. Positron emission tomography/magnetic resonance scans were acquired for 8 minutes per bed position (mpb). Positron emission tomography was extracted to reconstruct images with 2, 4, 6, and 8 mpb for each patient. Visual and quantitative approaches were used to assess image quality and lesion detectability for each image. For image quality, (a) 3 readers independently scored subjective image quality on a 4-point scale and (b) a region-of-interest approach was used to obtain a quantitative estimate of image quality in terms of noise. For lesion detectability, (a) the readers independently counted the number of hypermetabolic lesions and (b) signal-to-noise ratio and contrast-to-noise ratio were computed in a region-of-interest approach. Moreover, the mean and maximal standardized uptake value (SUV mean and SUV max, respectively) of the hypermetabolic lesions was compared across all acquisition times.
Results: For image quality, subjective image quality significantly declined from 8 to 2 mpb (P <; 0.05), with the exception of the difference between 6 and 8 mpb. Image noise increased with shorter imaging duration, ranging from 13% on average in the 8-mpb scans to 23% in the 2-mpb scans (differences were statistically significant for 2 vs 6 mpb, 2 vs 8 mpb, and 4 vs 8 mpb; P <; 0.05). For lesion detectability, 39 hypermetabolic lesions were identified by consensus. There was no difference in detected lesions across all acquisition times. Signal-to-noise ratio and contrast-to-noise ratio were constantly high and did not differ significantly across the acquisition times. The SUV mean and SUV max did not differ significantly across all acquisition times.
Conclusions: Positron emission tomography acquisition times on integrated PET/MRI do not need to exceed usual acquisition times on current PET/computed tomography scanners: Although the PET image quality suffers from short acquisition times, even a duration of 2 mpb permits sufficient lesion detection. Moreover, quantitative measures of tracer uptake are also reasonably precise at short acquisition times.
Similar articles
-
Integrated whole-body PET/MR hybrid imaging: clinical experience.Invest Radiol. 2013 May;48(5):280-9. doi: 10.1097/RLI.0b013e3182845a08. Invest Radiol. 2013. PMID: 23442775
-
Optimizing imaging protocols for overweight and obese patients: a lutetium orthosilicate PET/CT study.J Nucl Med. 2005 Apr;46(4):603-7. J Nucl Med. 2005. PMID: 15809482 Clinical Trial.
-
Evaluation of feasibility and image quality of 68Ga-DOTATOC positron emission tomography/magnetic resonance in comparison with positron emission tomography/computed tomography in patients with neuroendocrine tumors.Invest Radiol. 2013 May;48(5):263-72. doi: 10.1097/RLI.0b013e31828234d0. Invest Radiol. 2013. PMID: 23385399
-
Variability and uncertainty of 18F-FDG PET imaging protocols for assessing inflammation in atherosclerosis: suggestions for improvement.J Nucl Med. 2015 Apr;56(4):552-9. doi: 10.2967/jnumed.114.142596. Epub 2015 Feb 26. J Nucl Med. 2015. PMID: 25722452 Review.
-
PET-MRI fusion in head-and-neck oncology: current status and implications for hybrid PET/MRI.J Oral Maxillofac Surg. 2012 Feb;70(2):473-83. doi: 10.1016/j.joms.2011.02.120. Epub 2011 May 20. J Oral Maxillofac Surg. 2012. PMID: 21601338 Review.
Cited by
-
Optimized workflow and imaging protocols for whole-body oncologic PET/MRI.Jpn J Radiol. 2016 Nov;34(11):754-762. doi: 10.1007/s11604-016-0584-9. Epub 2016 Oct 6. Jpn J Radiol. 2016. PMID: 27714486
-
Locoregional tumour evaluation of squamous cell carcinoma in the head and neck area: a comparison between MRI, PET/CT and integrated PET/MRI.Eur J Nucl Med Mol Imaging. 2016 Jan;43(1):92-102. doi: 10.1007/s00259-015-3145-z. Epub 2015 Aug 6. Eur J Nucl Med Mol Imaging. 2016. PMID: 26243264
-
Optimized Whole-Body PET MRI Sequence Workflow in Pediatric Hodgkin Lymphoma Patients.J Nucl Med. 2023 Jan;64(1):96-101. doi: 10.2967/jnumed.122.264112. Epub 2022 Jul 14. J Nucl Med. 2023. PMID: 35835583 Free PMC article.
-
A novel rabbit fixator made of a thermoplastic mask for awake imaging experiments.Sci Rep. 2021 Jan 15;11(1):1546. doi: 10.1038/s41598-021-81358-6. Sci Rep. 2021. PMID: 33452449 Free PMC article.
-
Towards a fast PET/MRI protocol for breast cancer imaging: maintaining diagnostic confidence while reducing PET and MRI acquisition times.Eur Radiol. 2023 Sep;33(9):6179-6188. doi: 10.1007/s00330-023-09580-6. Epub 2023 Apr 12. Eur Radiol. 2023. PMID: 37045980 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical