A pilot trial of the iPad tablet computer as a portable device for visual acuity testing
- PMID: 23434538
- DOI: 10.1177/1357633X12474964
A pilot trial of the iPad tablet computer as a portable device for visual acuity testing
Abstract
We evaluated the accuracy of an app for the iPad tablet computer (Eye Chart Pro) as a portable method of visual acuity (VA) testing. A total of 120 consecutive patients (240 eyes) underwent visual acuity test with an iPad 2 and a conventional light-box chart. The logMAR VA results from the iPad were significantly higher than those from the light-box (P < 0.001). Bland-Altman analysis revealed a mean difference (bias) of 0.02 logMAR units between the VA results from the iPad chart and the light-box chart, with 95% limits of agreement of -0.14 to 0.19. Two groups of patients were defined: in Group 1 there were 182 eyes with VA better than 0.1 according to the light-box VA test. The median logMAR VA by the iPad was 0.54 and by the light-box chart it was 0.52; there was no significant difference between them (P = 0.69). In Group 2 there were 58 eyes with VA equal to or worse than 0.1 according to the light-box VA test. The median logMAR VA by the iPad was 1.26 and was 1.10 by the light box; the result from the iPad was significantly lower (P < 0.001). The Eye Chart Pro app installed on the iPad is reliable for VA testing only when the Snellen VA is better than 0.1 (20/200).