Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2013 Jan 1;7(1):170-8.
doi: 10.1177/193229681300700122.

Determination of hematocrit interference in blood samples derived from patients with different blood glucose concentrations

Affiliations

Determination of hematocrit interference in blood samples derived from patients with different blood glucose concentrations

Andreas Pfützner et al. J Diabetes Sci Technol. .

Abstract

Background: We performed a blood glucose meter hematocrit (HCT) interference test with lower sample manipulation requirements by using blood samples from patients with different blood glucose (BG) levels.

Methods: Blood from five patients with different BG levels (2.8, 5.6, 8.3, 13.9, 19.4 mmol/liter) was manipulated to contain five different HCT concentrations (35/40/45/50/55%). Each sample was measured three times in parallel with 14 BG testing devices (reference method: YSI 2300 STAT Plus™ Glucose Analyzer). The largest mean deviations in both directions from the reference method (normalized to 100% at 45% HCT) were added as a measure for hematocrit interference factor (HIF). A HIF >10% was considered to represent clinically relevant HCT interference.

Results: Few devices showed no clinically relevant HCT interference at high/low BG levels: BGStar® (7.2%, 7.3%), iBGStar® (9.0%, 8.6%), Contour® (10.0%, 4.6%), OneTouch® Verio™ 2 (10.0%, 5.2%), and GlucoMen® LX (7.2%, 5.1%). Other devices showed interference at one or both glucose ranges: ACCU-CHEK® Aviva (12.6%, 10.7%), Aviva Nano (7.2%, 10.5%), Breeze2 (3.6%, 30.2%), GlucoCard G+ (12.6%, 7.0%), OneTouch® Ultra®2 (12.6%, 25.6%), FreeStyle Freedom Lite® (9.0%, 11.0%), Precision Xceed (16.2%, 15.3%), and MediTouch® (19.8%, 28.0%). The deviations in all devices were less pronounced in the HCT range of 35-50%.

Conclusions: The results of this trial with less sample manipulation (HCT only) confirmed previous examinations with HCT and glucose manipulation. The same devices showed HCT stability as previously observed. Artificial sample manipulation may be less crucial than expected when evaluating HCT interference.

PubMed Disclaimer

Figures

Figure 1
Figure 1
Mean absolute deviation (in mg/dl) of the different BG meters from the reference method (YSI Stat 2300 Glucose Analyzer) in the samples with BG concentrations <6.5 mmol/liter (117 mg/dl) and with a HCT of 45%.
Figure 2
Figure 2
Mean absolute relative deviation (in %) of the different BG meters from the reference method (YSI Stat 2300 Glucose Analyzer) in the samples with BG concentrations >6.5 mmol/liter (117 mg/dl) and with a HCT of 45%.
Figure 3
Figure 3
The HIF (addition of the highest observed mean deviation over the entire glucose range above the 45% result to the highest absolute deviation below the 45% result) of the tested devices as calculated from the entire HCT range. Devices with an HIF <10% are considered to be stable against HCT interference.

References

    1. Ginsberg BH. Factors affecting blood glucose monitoring: Sources of error in measurement. J Diabetes Sci Technol. 2009;3(4):903–13. - PMC - PubMed
    1. Consensus statement on self-monitoring of blood glucose. Diabetes Care. 1987;10(1):95–9. - PubMed
    1. American Diabetes Association. Clinical practice recommendations 1996. Diabetes Care. 1996;19(Suppl 1):S1–118. - PubMed
    1. Pavlicek V, Garzoni D, Urech P, Brändle M. Inaccurate self-monitoring of blood glucose readings in patients on chronic ambulatory peritoneal dialysis with icodextrin. Exp Clin Endocrinol Diabetes. 2006;114(3):124–6. - PubMed
    1. Puntmann I, Wosniok W, Haeckel R. Comparison of several point-of-care testing (POCT) glucometers with an established laboratory procedure for the diagnosis of type 2 diabetes using the discordance rate: a new statistical approach. Clin Chem Lab Med. 2003;41(6):809–20. - PubMed

Publication types