. 2013 Mar;15(3):223-9.
doi: 10.1089/dia.2012.0292. Epub 2013 Feb 27.

Accuracy of continuous glucose monitoring during exercise in type 1 diabetes pregnancy

Affiliations

Accuracy of continuous glucose monitoring during exercise in type 1 diabetes pregnancy

Kavita Kumareswaran et al. Diabetes Technol Ther. 2013 Mar.

Abstract

Background: Performance of continuous glucose monitors (CGMs) may be lower when glucose levels are changing rapidly, such as occurs during physical activity. Our aim was to evaluate accuracy of a current-generation CGM during moderate-intensity exercise in type 1 diabetes (T1D) pregnancy.

Subjects and methods: As part of a study of 24-h closed-loop insulin delivery in 12 women with T1D (disease duration, 17.6 years; glycosylated hemoglobin, 6.4%) during pregnancy (gestation, 21 weeks), we evaluated the Freestyle Navigator(®) sensor (Abbott Diabetes Care, Alameda, CA) during afternoon (15:00-18:00 h) and morning (09:30-12:30 h) exercise (55 min of brisk walking on a treadmill followed by a 2-h recovery), compared with sedentary conditions (18:00-09:00 h). Plasma (reference) glucose, measured at regular 15-30-min intervals with the YSI Ltd. (Fleet, United Kingdom) model YSI 2300 analyzer, was used to assess CGM performance.

Results: Sensor accuracy, as indicated by the larger relative absolute difference (RAD) between paired sensor and reference glucose values, was lower during exercise compared with rest (median RAD, 11.8% vs. 18.4%; P<0.001). These differences remained significant when correcting for plasma glucose relative rate of change (P<0.001). Analysis by glucose range showed lower accuracy during hypoglycemia for both sedentary (median RAD, 24.4%) and exercise (median RAD, 32.1%) conditions. Using Clarke error grid analysis, 96% of CGM values were clinically safe under resting conditions compared with only 87% during exercise.

Conclusions: Compared with sedentary conditions, accuracy of the Freestyle Navigator CGM was lower during moderate-intensity exercise in pregnant women with T1D. This difference was particularly marked in hypoglycemia and could not be solely explained by the glucose rate of change associated with physical activity.

PubMed Disclaimer

Figures

FIG. 1.
FIG. 1.
Median relative absolute difference between paired sensor and plasma glucose values for sedentary and exercise periods, by glucose range.
FIG. 2.
FIG. 2.
Clarke error grid analysis showing paired sensor and plasma (reference) glucose values, during (A) sedentary and (B) exercise periods. There were 352 data pairs during the sedentary period and 138 during exercise.

References

    1. Ertl AC. Davis SN. Evidence for a vicious cycle of exercise and hypoglycemia in type 1 diabetes mellitus. Diabetes Metab Res Rev. 2004;20:124–130. - PubMed
    1. Wild D. von Maltzahn R. Brohan E. Christensen T. Clauson P. Gonder-Frederick L. A critical review of the literature on fear of hypoglycemia in diabetes: implications for diabetes management and patient education. Patient Educ Couns. 2007;68:10–15. - PubMed
    1. Zinman B. Ruderman N. Campaigne BN. Devlin JT. Schneider SH. Physical activity/exercise and diabetes mellitus. Diabetes Care. 2003;26(Suppl 1):S73–S77. - PubMed
    1. Devadoss M. Kennedy L. Herbold N. Endurance athletes and type 1 diabetes. Diabetes Educ. 2011;37:193–207. - PubMed
    1. Klonoff DC. Continuous glucose monitoring: roadmap for 21st century diabetes therapy. Diabetes Care. 2005;28:1231–1239. - PubMed

Publication types

MeSH terms

LinkOut - more resources