Value of semiquantitative analysis for clinical reporting of 123I-2-β-carbomethoxy-3β-(4-iodophenyl)-N-(3-fluoropropyl)nortropane SPECT studies
- PMID: 23492885
- DOI: 10.2967/jnumed.112.110106
Value of semiquantitative analysis for clinical reporting of 123I-2-β-carbomethoxy-3β-(4-iodophenyl)-N-(3-fluoropropyl)nortropane SPECT studies
Abstract
Clinical (123)I-2-β-carbomethoxy-3β-(4-iodophenyl)-N-(3-fluoropropyl)nortropane ((123)I-FP-CIT) SPECT studies are commonly performed and reported using visual evaluation of tracer binding, an inherently subjective method. Increased objectivity can potentially be obtained using semiquantitative analysis. In this study, we assessed whether semiquantitative analysis of (123)I-FP-CIT tracer binding created more reproducible clinical reporting. A secondary aim was to determine in what form semiquantitative data should be provided to the reporter.
Methods: Fifty-four patients referred for the assessment of nigrostriatal dopaminergic degeneration were scanned using SPECT/CT, followed by semiquantitative analysis calculating striatal binding ratios (SBRs) and caudate-to-putamen ratios (CPRs). Normal reference values were obtained using 131 healthy controls enrolled on a multicenter initiative backed by the European Association of Nuclear Medicine. A purely quantitative evaluation was first performed, with each striatum scored as normal or abnormal according to reference values. Three experienced nuclear medicine physicians then scored each striatum as normal or abnormal, also indicating cases perceived as difficult, using visual evaluation, visual evaluation in combination with SBR data, and visual evaluation in combination with SBR and CPR data. Intra- and interobserver agreement and agreement between observers and the purely quantitative evaluation were assessed using κ-statistics. The agreement between scan interpretation and clinical diagnosis was assessed for patients with a postscan clinical diagnosis available (n = 35).
Results: The physicians showed consistent reporting, with a good intraobserver agreement obtained for the visual interpretation (mean κ ± SD, 0.95 ± 0.029). Although visual interpretation of tracer binding gave good interobserver agreement (0.80 ± 0.045), this was improved as SBRs (0.86 ± 0.070) and CPRs (0.95 ± 0.040) were provided. The number of striata perceived as difficult to interpret decreased as semiquantitative data were provided (30 for the visual interpretation; 0 as SBR and CPR values were given). The agreement between physicians' interpretations and the purely quantitative evaluation showed that readers used the semiquantitative data to different extents, with a more experienced reader relying less on the semiquantitative data. Good agreement between scan interpretation and clinical diagnosis was seen.
Conclusion: A combined approach of visual assessment and semiquantitative analysis of tracer binding created more reproducible clinical reporting of (123)I-FP-CIT SPECT studies. Physicians should have access to both SBR and CPR data to minimize interobserver variability.
Keywords: 123I-FP-CIT; dopamine transporter; observer variability; semiquantification.
Similar articles
-
123I-FP-CIT SPECT imaging of the dopaminergic state. Visual assessment of dopaminergic degeneration patterns reflects quantitative 2D operator-dependent and 3D operator-independent techniques.Nuklearmedizin. 2012;51(6):244-51. doi: 10.3413/Nukmed-0449-11-12. Epub 2012 Apr 24. Nuklearmedizin. 2012. PMID: 22526237
-
Dopamine transporter loss visualized with FP-CIT SPECT in the differential diagnosis of dementia with Lewy bodies.Arch Neurol. 2004 Jun;61(6):919-25. doi: 10.1001/archneur.61.6.919. Arch Neurol. 2004. PMID: 15210531
-
Clinical testing of an optimized software solution for an automated, observer-independent evaluation of dopamine transporter SPECT studies.J Nucl Med. 2005 Jul;46(7):1109-18. J Nucl Med. 2005. PMID: 16000279 Clinical Trial.
-
A new era of clinical dopamine transporter imaging using 123I-FP-CIT.J Nucl Med Technol. 2012 Dec;40(4):222-8. doi: 10.2967/jnmt.112.111617. Epub 2012 Nov 16. J Nucl Med Technol. 2012. PMID: 23160562 Review.
-
Dopamine transporter imaging with [(123)I]FP-CIT SPECT: potential effects of drugs.Eur J Nucl Med Mol Imaging. 2008 Feb;35(2):424-38. doi: 10.1007/s00259-007-0621-0. Epub 2007 Oct 30. Eur J Nucl Med Mol Imaging. 2008. PMID: 17968545 Review.
Cited by
-
Investigation of Axial and Angular Sampling in Multi-Detector Pinhole-SPECT Brain Imaging.IEEE Trans Med Imaging. 2020 Dec;39(12):4209-4224. doi: 10.1109/TMI.2020.3015079. Epub 2020 Nov 30. IEEE Trans Med Imaging. 2020. PMID: 32763850 Free PMC article.
-
Improved Performance of a Multipinhole SPECT for DAT Imaging by Increasing Number of Pinholes at the Expense of Increased Multiplexing.IEEE Trans Radiat Plasma Med Sci. 2021 Nov;5(6):817-825. doi: 10.1109/trpms.2020.3035626. Epub 2020 Nov 3. IEEE Trans Radiat Plasma Med Sci. 2021. PMID: 34746540 Free PMC article.
-
Striatal Dopaminergic Depletion Pattern Reflects Pathological Brain Perfusion Changes in Lewy Body Diseases.Mol Imaging Biol. 2022 Dec;24(6):950-958. doi: 10.1007/s11307-022-01745-x. Epub 2022 Jun 14. Mol Imaging Biol. 2022. PMID: 35701723 Free PMC article.
-
Robust, fully automatic delineation of the head contour by stereotactical normalization for attenuation correction according to Chang in dopamine transporter scintigraphy.Eur Radiol. 2015 Sep;25(9):2709-17. doi: 10.1007/s00330-015-3667-6. Epub 2015 Feb 26. Eur Radiol. 2015. PMID: 25716941
-
Computer-aided diagnosis for (123I)FP-CIT imaging: impact on clinical reporting.EJNMMI Res. 2018 May 8;8(1):36. doi: 10.1186/s13550-018-0393-5. EJNMMI Res. 2018. PMID: 29740722 Free PMC article.
Publication types
MeSH terms
Substances
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources