. 2013 May;18(5):490-5.
doi: 10.3171/2013.2.SPINE12813. Epub 2013 Mar 15.

Stereotactic navigation with the O-arm for placement of S-2 alar iliac screws in pelvic lumbar fixation

Affiliations

Stereotactic navigation with the O-arm for placement of S-2 alar iliac screws in pelvic lumbar fixation

Wilson Z Ray et al. J Neurosurg Spine. 2013 May.

Abstract

Object: Pelvic fixation is a crucial adjunct to many lumbar fusions to avoid L5-S1 pseudarthrosis. It is useful for treatment of kyphoscoliosis, high-grade spondylolisthesis, L5-S1 pseudarthrosis, sacral tumors, lumbosacral dislocations, and osteomyelitis. The most popular method, iliac fixation, has drawbacks including hardware prominence, extensive muscle dissection, and the need for connection devices. S-2 alar iliac fixation provides a useful primary or salvage alternative. The authors describe their techniques for using stereotactic navigation for screw placement.

Methods: The O-arm Surgical Imaging System allowed for CT-quality multiplanar reconstructions of the pelvis, and registration to a StealthStation Treon provided intraoperative guidance. The authors describe their technique for performing computer-assisted S-2 alar iliac fixation for various indications in 18 patients during an 18-month period.

Results: All patients underwent successful bilateral placement of screws 80-100 mm in length. All placements were confirmed with a second multiplanar reconstruction. One screw was moved because of apparent anterior breach of the ilium. There were no immediate neurological or vascular complications due to screw placement. The screw length required additional instruments including a longer pedicle finder and tap.

Conclusions: Stereotactic guidance to navigate the placement of distal pelvic fixation with bilateral S-2 alar iliac fixation can be safely performed in patients with a variety of pathological conditions. Crossing the sacroiliac joint, choosing trajectory, and ensuring adequate screw length can all be enhanced with 3D image guidance. Long-term outcome studies are underway, specifically evaluating the sacroiliac joint.

PubMed Disclaimer

Comment in

  • Letter to the editor: The O-arm revolution in spine surgery.
    Mattei TA, Fassett DR. Mattei TA, et al. J Neurosurg Spine. 2013 Nov;19(5):644-7. doi: 10.3171/2013.5.SPINE13474. Epub 2013 Sep 20. J Neurosurg Spine. 2013. PMID: 24053374 No abstract available.
  • Response.
    Ray W, Dailey AT. Ray W, et al. J Neurosurg Spine. 2013 Nov;19(5):647. J Neurosurg Spine. 2013. PMID: 24344464 No abstract available.