Purpose: To describe the first use of the commercially available Calypso 4D Localization System in the lung.
Methods and materials: Under an institutional review board-approved protocol and an investigational device exemption from the US Food and Drug Administration, the Calypso system was used with nonclinical methods to acquire real-time 4-dimensional lung tumor tracks for 7 lung cancer patients. The aims of the study were to investigate (1) the potential for bronchoscopic implantation; (2) the stability of smooth-surface beacon transponders (transponders) after implantation; and (3) the ability to acquire tracking information within the lung. Electromagnetic tracking was not used for any clinical decision making and could only be performed before any radiation delivery in a research setting. All motion tracks for each patient were reviewed, and values of the average displacement, amplitude of motion, period, and associated correlation to a sinusoidal model (R(2)) were tabulated for all 42 tracks.
Results: For all 7 patients at least 1 transponder was successfully implanted. To assist in securing the transponder at the tumor site, it was necessary to implant a secondary fiducial for most transponders owing to the transponder's smooth surface. For 3 patients, insertion into the lung proved difficult, with only 1 transponder remaining fixed during implantation. One patient developed a pneumothorax after implantation of the secondary fiducial. Once implanted, 13 of 14 transponders remained stable within the lung and were successfully tracked with the tracking system.
Conclusions: Our initial experience with electromagnetic guidance within the lung demonstrates that transponder implantation and tracking is achievable though not clinically available. This research investigation proved that lung tumor motion exhibits large variations from fraction to fraction within a single patient and that improvements to both transponder and tracking system are still necessary to create a clinical daily-use system to assist with actual lung radiation therapy.
Copyright © 2013 Elsevier Inc. All rights reserved.
Similar articles
-
Anchored Transponder Guided Lung Radiation Therapy.Pract Radiat Oncol. 2020 Jan-Feb;10(1):e37-e44. doi: 10.1016/j.prro.2019.08.009. Epub 2019 Aug 31. Pract Radiat Oncol. 2020. PMID: 31479771 Clinical Trial.
-
Bronchoscopic implantation of a novel wireless electromagnetic transponder in the canine lung: a feasibility study.Int J Radiat Oncol Biol Phys. 2008 Sep 1;72(1):93-8. doi: 10.1016/j.ijrobp.2007.12.055. Epub 2008 Apr 18. Int J Radiat Oncol Biol Phys. 2008. PMID: 18394817
-
Target localization and real-time tracking using the Calypso 4D localization system in patients with localized prostate cancer.Int J Radiat Oncol Biol Phys. 2006 Jun 1;65(2):528-34. doi: 10.1016/j.ijrobp.2006.01.050. Int J Radiat Oncol Biol Phys. 2006. PMID: 16690435
-
Longitudinal assessment of anchored transponder migration following lung stereotactic body radiation therapy.J Appl Clin Med Phys. 2019 Jan;20(1):17-22. doi: 10.1002/acm2.12454. Epub 2018 Nov 1. J Appl Clin Med Phys. 2019. PMID: 30387242 Free PMC article.
-
Feasibility of electromagnetic transponder use to monitor inter- and intrafractional motion in locally advanced pancreatic cancer patients.Int J Radiat Oncol Biol Phys. 2012 Jun 1;83(2):566-73. doi: 10.1016/j.ijrobp.2011.07.025. Epub 2011 Nov 16. Int J Radiat Oncol Biol Phys. 2012. PMID: 22099029
Cited by
-
Dosimetric Study of Deep Learning-Guided ITV Prediction in Cone-beam CT for Lung Stereotactic Body Radiotherapy.Front Public Health. 2022 Mar 22;10:860135. doi: 10.3389/fpubh.2022.860135. eCollection 2022. Front Public Health. 2022. PMID: 35392465 Free PMC article.
-
Influence of respiratory motion management technique on radiation pneumonitis risk with robotic stereotactic body radiation therapy.J Appl Clin Med Phys. 2018 Jul;19(4):48-57. doi: 10.1002/acm2.12338. Epub 2018 Apr 26. J Appl Clin Med Phys. 2018. PMID: 29700954 Free PMC article.
-
Commissioning of a motion system to investigate dosimetric consequences due to variability of respiratory waveforms.J Appl Clin Med Phys. 2016 Jan 8;17(1):283-292. doi: 10.1120/jacmp.v17i1.5921. J Appl Clin Med Phys. 2016. PMID: 26894366 Free PMC article.
-
Clinical experience with lung-specific electromagnetic transponders for real-time tumor tracking in lung stereotactic body radiotherapy.Phys Imaging Radiat Oncol. 2019 Nov 28;12:30-37. doi: 10.1016/j.phro.2019.11.002. eCollection 2019 Oct. Phys Imaging Radiat Oncol. 2019. PMID: 33458292 Free PMC article.
-
MArkerless image Guidance using Intrafraction Kilovoltage x-ray imaging (MAGIK): study protocol for a phase I interventional study for lung cancer radiotherapy.BMJ Open. 2022 Jan 20;12(1):e057135. doi: 10.1136/bmjopen-2021-057135. BMJ Open. 2022. PMID: 35058267 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical