Prospective-triggered sequential dual-source end-systolic coronary CT angiography for patients with atrial fibrillation: a feasibility study
- PMID: 23545461
- DOI: 10.1016/j.jcct.2013.02.002
Prospective-triggered sequential dual-source end-systolic coronary CT angiography for patients with atrial fibrillation: a feasibility study
Abstract
Background: Obtaining diagnostic coronary CT angiography with low radiation exposure in patients with irregular heart rhythms such as atrial fibrillation (AF) remains challenging.
Objective: We evaluated image quality and inter-reader variability with the use of prospective electrocardiographic (ECG)-triggered sequential dual-source acquisition at end systole for coronary artery disease (CAD) evaluation in patients with AF.
Methods: Thirty consecutive patients with AF who underwent prospective ECG-triggered sequential dual-source acquisition were evaluated. Images were reconstructed every 50 milliseconds from 250 to 400 milliseconds after the R wave. Two independent, blinded readers evaluated the coronaries for image quality on a 5-point scale (worst to best) and stenosis on 5-point semiquantitative (none to severe) and binary scales (>50% or <50%). Diagnostic image quality was graded for each reconstruction.
Results: Eleven patients (37%) had significant (≥50% stenosis) CAD. Average heart rate was 82 ± 20 beats/min and variability range was 71 ± 22 beats/min. Mean effective radiation dose was 6.5 ± 2.4 mSv. Diagnostic image quality was noted in 97.9% of 304 coronary segments with median image quality of 3.0. The 300-millisecond reconstruction phase provided the highest image quality; 70% of patients showed diagnostic image quality. Combination of all phases (250-400 milliseconds) performed significantly better than single or other phase combinations (P < 0.0005 for all comparisons). Inter-reader variability for stenosis detection was excellent, with 98.4% concordance by using a binary scale (50% stenosis cutoff).
Conclusions: Prospective ECG-triggered sequential dual-source CT acquisition with the use of end-systolic acquisition provides diagnostic image quality with potentially low radiation doses for evaluation of CAD in patients with AF. Use of multiple end-systolic phases over a 150-millisecond window improves diagnostic image quality.
Copyright © 2013 Society of Cardiovascular Computed Tomography. Published by Elsevier Inc. All rights reserved.
Similar articles
-
Prospectively ECG-Triggered Sequential Dual-Source Coronary CT Angiography in Patients with Atrial Fibrillation: Influence of Heart Rate on Image Quality and Evaluation of Diagnostic Accuracy.PLoS One. 2015 Jul 29;10(7):e0134194. doi: 10.1371/journal.pone.0134194. eCollection 2015. PLoS One. 2015. PMID: 26221952 Free PMC article.
-
Noninvasive coronary angiography using dual-source computed tomography in patients with atrial fibrillation.Invest Radiol. 2009 Mar;44(3):159-67. doi: 10.1097/RLI.0b013e3181948b05. Invest Radiol. 2009. PMID: 19151607
-
Dual-source CT coronary angiography in patients with atrial fibrillation: comparison with single-source CT.Eur J Radiol. 2008 Dec;68(3):434-41. doi: 10.1016/j.ejrad.2008.09.011. Epub 2008 Nov 6. Eur J Radiol. 2008. PMID: 18993003
-
Current and Novel Imaging Techniques in Coronary CT.Radiographics. 2015 Jul-Aug;35(4):991-1010. doi: 10.1148/rg.2015140181. Epub 2015 Jun 5. Radiographics. 2015. PMID: 26046942 Review.
-
Cardiovascular imaging in pediatric patients using dual source CT.J Cardiovasc Comput Tomogr. 2016 Jan-Feb;10(1):13-21. doi: 10.1016/j.jcct.2015.10.003. Epub 2015 Oct 19. J Cardiovasc Comput Tomogr. 2016. PMID: 26524989 Review.
Cited by
-
Diagnostic quality of dual-source coronary CT examinations performed without heart rate control: importance of obesity and heart rate on image quality.J Comput Assist Tomogr. 2014 Nov-Dec;38(6):949-55. doi: 10.1097/RCT.0000000000000135. J Comput Assist Tomogr. 2014. PMID: 25032805 Free PMC article.
-
Prospectively ECG-Triggered Sequential Dual-Source Coronary CT Angiography in Patients with Atrial Fibrillation: Influence of Heart Rate on Image Quality and Evaluation of Diagnostic Accuracy.PLoS One. 2015 Jul 29;10(7):e0134194. doi: 10.1371/journal.pone.0134194. eCollection 2015. PLoS One. 2015. PMID: 26221952 Free PMC article.
-
Coronary CT angiography in patients with atrial fibrillation: Standard-dose and low-dose imaging with a high-resolution whole-heart CT scanner.Eur Radiol. 2018 Aug;28(8):3432-3440. doi: 10.1007/s00330-017-5282-1. Epub 2018 Feb 9. Eur Radiol. 2018. PMID: 29426987
-
Role of atrial wall thickness in wave-dynamics of atrial fibrillation.PLoS One. 2017 Aug 21;12(8):e0182174. doi: 10.1371/journal.pone.0182174. eCollection 2017. PLoS One. 2017. PMID: 28827810 Free PMC article.
-
Diagnosis of coronary artery disease in patients with atrial fibrillation using low tube voltage coronary CT angiography with isotonic low-concentration contrast agent.Int J Cardiovasc Imaging. 2019 Dec;35(12):2239-2248. doi: 10.1007/s10554-019-01678-7. Epub 2019 Jul 30. Int J Cardiovasc Imaging. 2019. PMID: 31363878
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Miscellaneous