[Opacification of an intraocular lens: calcification of hydrophilic intraocular lenses after gas tamponade of the anterior chamber]
- PMID: 23552856
- DOI: 10.1007/s00347-013-2793-2
[Opacification of an intraocular lens: calcification of hydrophilic intraocular lenses after gas tamponade of the anterior chamber]
Abstract
A patient with endothelial dystrophy was treated with Descemet stripping automated endothelial keratoplasty (DSAEK) combined with cataract extraction and implantation of a hydrophilic intraocular lens (IOL, Lentis-L312, Oculentis) but visual acuity dropped from 0.15 logMAR to 0.52 logMAR 18 months later due to calcification of the IOL. With new methods of lamellar corneal transplantation being used more frequently the number of necessary anterior chamber tamponades with air/gas are increasing. In cataract cases in which a gas tamponade and transplantation might be necessary later on (cornea guttata), hydrophilic IOLs should be avoided.
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources