Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Comparative Study
. 2013;20(2):161-9.
doi: 10.5603/CJ.2013.0029.

Broken leads with proximal endings in the cardiovascular system: serious consequences and extraction difficulties

Affiliations
Free article
Comparative Study

Broken leads with proximal endings in the cardiovascular system: serious consequences and extraction difficulties

Andrzej Kutarski et al. Cardiol J. 2013.
Free article

Abstract

Background: Retrospective analysis of effectiveness, technical problems, and complications of transvenous extraction of leads with the free endings migrated to the cardiovascular system (CVS).

Methods: A 5-year-old database of transvenous lead extraction (TLE) procedures comprising 906 patients with 1563 leads being removed was analyzed. TLE procedures of leads migrated in the CVS were compared with TLE procedures of leads with their proximal ends accessible in the pacemaker/implantable cardioverter-defibrillator (PM/ICD) pocket.

Results: In our material, the phenomenon of leads migration occurred in 5% of patients referred for TLE and affected most frequently unipolar and atrial leads. The presence of migrating leads was associated with local venous occlusion in 64% of patients. Removal of migrating leads required other techniques than extraction of leads with their proximal ends accessible in the PM/ICD pocket. More than 95% of migrating leads were extracted transvenously, but procedures were significantly longer. The presence of other leads made extraction of migrated leads even more complicated. Effectiveness and complication rates for removal of migrated leads and leads accessible in the PM/ICD pocket were similar.

Conclusions: We postulate that every lead migrating in the CVS should be considered for TLE. However, this extraction is technically more difficult and challenging than extraction of leads accessible in the PM/ICD pocket.