Accuracy of the CNAP™ monitor, a noninvasive continuous blood pressure device, in providing beat-to-beat blood pressure readings in pediatric patients weighing 20-40 kilograms
- PMID: 23565770
- DOI: 10.1111/pan.12173
Accuracy of the CNAP™ monitor, a noninvasive continuous blood pressure device, in providing beat-to-beat blood pressure readings in pediatric patients weighing 20-40 kilograms
Abstract
Background: During perioperative care, the continuous measurement of blood pressure (BP) provides superior physiologic monitoring to intermittent techniques. However, such monitoring requires placement of an intraarterial catheter, which may be time-consuming or associated with adverse events and technical difficulty. A noninvasive, continuous BP monitoring device has been studied in the adult population. This study prospectively assesses its accuracy in pediatric patients, weighing 20-40 kg.
Methods: The technology evaluated is the CNAP™ Monitor 500, developed by CNSystems AG (Graz, Austria). The study cohort included pediatric patients weighing between 20 and 40 kg, scheduled for surgery for which arterial line (AL) placement was planned. Systolic (sBP), diastolic (dBP), and mean arterial (MAP) blood pressure readings were captured from the AL and the CNAP™ device every minute during anesthetic care.
Results: The study cohort consisted of 20 patients (11 weighing between 30 and 40 kg and 9 weighing between 20 and 29.9 kg) with a mean age of 9.8 ± 3.4 years (range, 6-16 years) and weight of 29.8 ± 6.1 kg (range, 20.9-38.7 kg). There were a total of 1076 pairs each of sBP, dBP, and MAP values in the 20-29.9 kg group. The absolute difference between the sBP, dBP, and MAP was 9.8 ± 8.5, 6.8 ± 5.3, and 6.7 ± 6.2 mmHg, respectively. The correlation coefficient between the AL and the CNAP™ device was 0.48, 0.60, and 0.64 for the sBP, dBP, and MAP, respectively. The CNAP™ values (sBP, dBP, MAP) were ≤5 mmHg from the AL values in 38.6%, 48.5%, and 55.0% of the values, respectively. In the 30-40 kg group, there were a total of 2737 pairs of sBP, dBP, and MAP values. The absolute difference between the sBP, dBP, and MAP was 11.5 ± 9.3, 7.5 ± 5.3, and 7.9 ± 6.6 mmHg, respectively. The correlation coefficient between the arterial cannula and the CNAP™ device was 0.48, 0.45, and 0.51 for the sBP, dBP, and MAP, respectively. CNAP™ readings were ≤5 mmHg from the AL values (sBP, dBP, MAP) in 29.0%, 41.9%, and 40.5% of the values, respectively.
Conclusion: Although some variation in its accuracy was noted, the CNAP™ device provides a noninvasive and continuous blood pressure reading which appears to be within clinically useful limits. It may be that modification of the finger cuffs is needed to improve its absolute accuracy as our clinical experience demonstrated that achieving an effective fit with the cuffs was at times difficult.
Keywords: arterial blood pressure; blood pressure monitors; pediatrics.
© 2013 John Wiley & Sons Ltd.
Similar articles
-
Accuracy of the CNAP™ monitor, a noninvasive continuous blood pressure device, in providing beat-to-beat blood pressure measurements during bariatric surgery in severely obese adolescents and young adults.J Anesth. 2014 Dec;28(6):861-5. doi: 10.1007/s00540-014-1835-5. Epub 2014 May 1. J Anesth. 2014. PMID: 24789660
-
Accuracy of the CNAP monitor, a noninvasive continuous blood pressure device, in providing beat-to-beat blood pressure readings in the prone position.J Clin Anesth. 2013 Jun;25(4):309-13. doi: 10.1016/j.jclinane.2013.01.010. Epub 2013 May 17. J Clin Anesth. 2013. PMID: 23685101
-
A comparison of a continuous noninvasive arterial pressure (CNAP™) monitor with an invasive arterial blood pressure monitor in the cardiac surgical ICU.Ann Card Anaesth. 2012 Jul-Sep;15(3):180-4. doi: 10.4103/0971-9784.97973. Ann Card Anaesth. 2012. PMID: 22772511
-
Update on Finger-Application-Type Noninvasive Continuous Hemodynamic Monitors (CNAP and ccNexfin): Physical Principles, Validation, and Clinical Use.Semin Cardiothorac Vasc Anesth. 2017 Dec;21(4):321-329. doi: 10.1177/1089253217708620. Epub 2017 May 10. Semin Cardiothorac Vasc Anesth. 2017. PMID: 28490233 Review.
-
Impacts of measurement protocols on blood pressure tracking from childhood into adulthood: a metaregression analysis.Hypertension. 2008 Mar;51(3):642-9. doi: 10.1161/HYPERTENSIONAHA.107.102145. Epub 2008 Jan 22. Hypertension. 2008. PMID: 18212267 Review.
Cited by
-
Accuracy of the CNAP™ monitor, a noninvasive continuous blood pressure device, in providing beat-to-beat blood pressure measurements during bariatric surgery in severely obese adolescents and young adults.J Anesth. 2014 Dec;28(6):861-5. doi: 10.1007/s00540-014-1835-5. Epub 2014 May 1. J Anesth. 2014. PMID: 24789660
-
Wearable biosensors for pediatric hospitals: a scoping review.Pediatr Res. 2025 Jul;98(1):90-99. doi: 10.1038/s41390-024-03693-4. Epub 2024 Nov 7. Pediatr Res. 2025. PMID: 39511444
-
An Evaluation of the NICCI Monitor in Providing Continuous, Noninvasive Blood Pressure Readings in Children During Intraoperative Anesthetic Care.J Clin Med Res. 2022 Apr;14(4):158-164. doi: 10.14740/jocmr4723. Epub 2022 Apr 30. J Clin Med Res. 2022. PMID: 35573930 Free PMC article.
-
Using what you get: dynamic physiologic signatures of critical illness.Crit Care Clin. 2015 Jan;31(1):133-64. doi: 10.1016/j.ccc.2014.08.007. Crit Care Clin. 2015. PMID: 25435482 Free PMC article. Review.
-
Measurement of Blood Pressure by Ultrasound-The Applicability of Devices, Algorithms and a View in Local Hemodynamics.Diagnostics (Basel). 2021 Dec 2;11(12):2255. doi: 10.3390/diagnostics11122255. Diagnostics (Basel). 2021. PMID: 34943492 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources