Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Clinical Trial
. 2013 Mar 1;7(2):356-61.
doi: 10.1177/193229681300700211.

Acceptability and utility of the mySentry remote glucose monitoring system

Affiliations
Clinical Trial

Acceptability and utility of the mySentry remote glucose monitoring system

Kevin Kaiserman et al. J Diabetes Sci Technol. .

Abstract

Background: The mySentry system (Medtronic Inc.) is the first to amplify and relay continuous glucose monitoring (CGM) and insulin pump data to a remote site within the house. Its usability and acceptability were evaluated in families having a child with type 1 diabetes.

Methods: Each enrolled family included a child (age 7-17 years) who used a Paradigm REAL-Time Revel sensor-augmented insulin pump (Medtronic). After a 1-week run-in phase, families set up and used the mySentry system for a 3-week study phase. Opinion surveys were completed by parents, and pump and CGM data were collected and analyzed retrospectively. No formal hypothesis testing was performed, and the study was not powered to detect changes in nocturnal glycemia.

Results: Thirty-five families completed the study. Enrolled children (61.1% female) had a mean (± standard deviation) age of 11.9 ± 2.70 years and a mean age at initiation of pump therapy of 7.1 ± 3.19 years. Baseline survey results indicated that most parents were fearful of their unawareness of their children's nocturnal glucose excursions. The mySentry system met the predefined acceptability criteria for general experience, product usability, and training materials. There were no unanticipated device-related adverse effects. Among children who experienced nocturnal hypo- or hyperglycemic episodes in both phases of the study, there was a trend toward less frequent and less prolonged episodes during mySentry use.

Conclusion: The mySentry system met all predefined criteria for acceptability and did not demonstrate safety issues. Alerting parents to abnormal glucose values or trends may attenuate nocturnal hypoglycemia and hyperglycemia by prompting appropriate and timely intervention.

PubMed Disclaimer

Figures

Figure 1
Figure 1
The outpost (A) and monitor (B) of the mySentry system.

Comment in

References

    1. Davis EA, Keating B, Byrne GC, Russell M, Jones TW. Hypoglycemia: incidence and clinical predictors in a large population-based sample of children and adolescents with IDDM. Diabetes Care. 1997;20(1):22–5. - PubMed
    1. Elleri D, Acerini CL, Allen JM, Hayes J, Pesterfield C, Wilinska ME, Dunger DB, Hovorka R. Parental attitudes towards overnight closed-loop glucose control in children with type 1 diabetes. Diabetes Technol Ther. 2010;12(1):35–9. - PubMed
    1. Tanenberg RJ, Newton CA, Drake AJ. Confirmation of hypoglycemia in the “dead-in-bed” syndrome, as captured by a retrospective continuous glucose monitoring system. Endocr Pract. 2010;16(2):244–8. - PubMed
    1. Nordfeldt S, Ludvigsson J. Fear and other disturbances of severe hypoglycaemia in children and adolescents with type 1 diabetes mellitus. J Pediatr Endocrinol Metab. 2005;18(1):83–91. - PubMed
    1. Matyka KA. Nocturnal hypoglycaemia in children: the effects on cognitive function. Diabetes Nutr Metab. 2002;15(6):390–4. - PubMed

Publication types

MeSH terms