Optimizing bone screw pullout force
- PMID: 2358933
- DOI: 10.1097/00005131-199004020-00012
Optimizing bone screw pullout force
Abstract
Stability of internal fixation by plate and screw depends on the interface of bone and screw threads. Bone-screw thread design was tested in a synthetic bone model to determine maximum bone-screw pullout force. The model was confirmed to provide reproducible results in a clinically relevant range. Consistent material properties were reflected in the relatively small standard deviations in pullout force, which were less than 10% of mean values. As expected, major diameter was an important determinant of pullout force in a roughly linear manner. Pitch was important with a finer thread giving greater purchase. Minor diameter and the ratio of major to minor diameter had a small but significant effect on pullout force. This study was significant for establishing a good bone-screw testing model and suggesting important thread parameters in selecting or manufacturing a bone screw to optimize its holding power.
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources
Research Materials