Transcatheter interatrial septal defect closure in a large cohort: midterm follow-up results
- PMID: 23601507
- DOI: 10.1111/chd.12057
Transcatheter interatrial septal defect closure in a large cohort: midterm follow-up results
Abstract
Objectives: We evaluated immediate and midterm results of transcatheter closure of atrial septal defects (ASDs) and patent foramen ovale (PFO) using various closure devices.
Materials and methods: The study included four hundred fourteen patients (one hundred eighty-two men, two hundred thirty-two women; mean age 39 ± 12.3 years; range 17-67 years) who underwent transcatheter closure of secundum ASD (n = 193) or PFO (n = 221). All the patients were evaluated by transthoracic echocardiography and transesophageal echocardiography before the procedure. Transcatheter closure was performed by using Amplatzer (n = 184), Occlutech Figulla (n = 209), or BioSTAR (n = 21) devices. Closure of ASDs was performed under general anesthesia with transesophageal echocardiography guidance, and closure of PFOs was performed under local anesthesia with transthoracic echocardiography guidance. Follow-up controls were at 1, 6, and 12 months and annually thereafter. The median follow-up periods of ASD and PFO patients were 43 and 30 months.
Results: The mean device size was 19.3 ± 6.2 mm for ASD patients and 24.6 ± 2.6 mm for PFO patients. The mean procedural and fluoroscopy times were 22.3 ± 4.7 and 4.1 ± 1.9 minutes for ASD closure and 12.4 ± 3.2 and 3.1 ± 1.2 minutes for PFO closure, respectively. Procedural device embolization occurred in only two patients (0.48%). During follow-up, recurrent embolic events occurred in four patients (1.8%) after PFO closure, and no residual shunts were seen after ASD closure. Device thrombosis developed in two ASD patients during the procedure and in one PFO patient at 12th month of the follow-up (0.72%).
Conclusion: Transcatheter closure of PFOs and secundum-type ASDs using the Amplatzer, Occlutech Figulla, and BioSTAR devices is an efficacious and safe therapeutic option.
Keywords: Atrial Septal Defect; Occluder Device; Patent Foramen Ovale.
© 2013 Wiley Periodicals, Inc.
Similar articles
-
[Percutaneous closure of interatrial septal defects: mid-term follow-up results].Turk Kardiyol Dern Ars. 2011 Jul;39(5):385-95. doi: 10.5543/tkda.2011.01474. Turk Kardiyol Dern Ars. 2011. PMID: 21743262 Turkish.
-
The Flatstent versus the conventional umbrella devices in the percutaneous closure of patent foramen ovale.Catheter Cardiovasc Interv. 2015 May;85(6):1058-65. doi: 10.1002/ccd.25750. Epub 2014 Nov 27. Catheter Cardiovasc Interv. 2015. PMID: 25413379
-
Impact of Percutaneous Closure of Interatrial Shunts on Migraine Attacks: Single-Operator Series and Review of the Literature.Rev Recent Clin Trials. 2017;12(2):129-138. doi: 10.2174/1574887112666170328124939. Rev Recent Clin Trials. 2017. PMID: 28356032 Review.
-
The Occlutech Figulla PFO and ASD occluder: a new nitinol wire mesh device for closure of atrial septal defects.J Invasive Cardiol. 2010 Apr;22(4):182-7. J Invasive Cardiol. 2010. PMID: 20351390
-
Short and long term complications of device closure of atrial septal defect and patent foramen ovale: meta-analysis of 28,142 patients from 203 studies.Catheter Cardiovasc Interv. 2013 Dec 1;82(7):1123-38. doi: 10.1002/ccd.24875. Epub 2013 Aug 31. Catheter Cardiovasc Interv. 2013. PMID: 23412921 Review.
Cited by
-
The influence of deficient retro-aortic rim on technical success and early adverse events following device closure of secundum atrial septal defects: An Analysis of the IMPACT Registry®.Catheter Cardiovasc Interv. 2017 Jan;89(1):102-111. doi: 10.1002/ccd.26585. Epub 2016 May 18. Catheter Cardiovasc Interv. 2017. PMID: 27189502 Free PMC article.
-
Transcatheter device closure of atrial septal defects guided completely by transthoracic echocardiography: A single cardiac center experience with 152 cases.Anatol J Cardiol. 2018 Dec;20(6):330-335. doi: 10.14744/AnatolJCardiol.2018.90502. Anatol J Cardiol. 2018. PMID: 30504733 Free PMC article.
-
Percutaneous Atrial Septal Defect Closure Using the Occlutech Figulla Device in Adults: More than 800 Patient-Years of Follow-Up.J Interv Cardiol. 2020 Feb 14;2020:7136802. doi: 10.1155/2020/7136802. eCollection 2020. J Interv Cardiol. 2020. PMID: 32140088 Free PMC article.
-
Comparison of Figulla Flex® and Amplatzer™ devices for atrial septal defect closure: A meta-analysis.Cardiol J. 2020;27(5):524-532. doi: 10.5603/CJ.a2020.0058. Epub 2020 Apr 24. Cardiol J. 2020. PMID: 32329042 Free PMC article.
-
Microtransesophageal Echocardiographic Guidance during Percutaneous Interatrial Septal Closure without General Anaesthesia.J Interv Cardiol. 2020 Sep 7;2020:1462140. doi: 10.1155/2020/1462140. eCollection 2020. J Interv Cardiol. 2020. PMID: 32982607 Free PMC article.
MeSH terms
Supplementary concepts
LinkOut - more resources
Full Text Sources
Other Literature Sources