. 2013;26(2):141-8.
doi: 10.3233/BMR-2012-00358.

Measurement of cervical range of motion (CROM) by electronic CROM goniometer: a test of reliability and validity

Affiliations

Measurement of cervical range of motion (CROM) by electronic CROM goniometer: a test of reliability and validity

Ellis Yuk Hung Law et al. J Back Musculoskelet Rehabil. 2013.

Abstract

Purpose: To investigate the reliability and validity of the Electronic Cervical Range of Motion (CROM) Goniometer in measurement of cervical spine mobility in adults with and without neck pain.

Methods: A cross-sectional reliability study was conducted on 54 subjects (26 neck pain and 26 non-neck pain) aged from 20-70 years old. The Numerical Pain Rating Scale and Chinese version of Northwick Park Neck Pain Questionnaire were used to assess neck pain severity and disability respectively. The CROM was measured in sitting position except left to right rotation was measured in supine lying. All the cervical active movements were measured by using the Electronic CROM Goniometer from ARCON TM Functional Capacity Evaluation (FCE) systems.

Results: The intra-tester and inter-tester reliability were high in both normal and chronic neck pain groups with ICC coefficients ranged from 0.75-*0.92. There was significant difference in the total CROM between the normal (374.7°) and chronic neck pain group (292.6°).

Conclusion: The ACRON cervical goniometer was found to be reliable for measuring cervical mobility in 3 planes for both normal and patient subjects. Construct validity of the goniometer was supported as the test's result documented significant difference in CROM between the control and the neck pain groups.