Cerebral TOF angiography at 7T: Impact of B1 (+) shimming with a 16-channel transceiver array
- PMID: 23640915
- PMCID: PMC4097948
- DOI: 10.1002/mrm.24749
Cerebral TOF angiography at 7T: Impact of B1 (+) shimming with a 16-channel transceiver array
Abstract
Purpose: Time-of-flight (TOF) MR imaging is clinically among the most common cerebral noncontrast enhanced MR angiography techniques allowing for high spatial resolution. As shown by several groups TOF contrast significantly improves at ultrahigh field of B0 = 7T, however, spatially varying transmit B1 (B1 (+)) fields at 7T reduce TOF contrast uniformity, typically resulting in suboptimal contrast and reduced vessel conspicuity in the brain periphery.
Methods: Using a 16-channel B1 (+) shimming system, we compare different dynamically applied B1 (+) phase shimming approaches on the radiofrequency excitation to improve contrast homogeneity for a (0.5 mm)(3) resolution multislab TOF acquisition. In addition, B1 (+) shimming applied on the venous saturation pulse was investigated to improve venous suppression, subcutaneous fat signal reduction and enhanced background suppression originating from MT effect.
Results: B1 (+) excitation homogeneity was improved by a factor 2.2-2.6 on average depending on the shimming approach, compared to a standard CP-like phase setting, leading to improved vessel conspicuity particularly in the periphery. Stronger saturation, higher fat suppression and improved background suppression were observed when dynamically applying B1 (+) shimming on the venous saturation pulse.
Conclusion: B1+ shimming can significantly improve high resolution TOF vascular investigations at ultrahigh field, holding strong promise for non contrast-enhanced clinical applications.
Copyright © 2013 Wiley Periodicals, Inc.
Figures
References
-
- Heverhagen JT, Bourekas E, Sammet S, Knopp MV, Schmalbrock P. Time-of-flight magnetic resonance angiography at 7 Tesla. Invest Radiol. 2008;43(8):568–573. - PubMed
-
- Kang CK, Park CW, Han JY, Kim SH, Park CA, Kim KN, Hong SM, Kim YB, Lee KH, Cho ZH. Imaging and analysis of lenticulostriate arteries using 7. 0-Tesla magnetic resonance angiography. Magn Reson Med. 2009;61(1):136–144. - PubMed
-
- Zwanenburg JJ, Hendrikse J, Takahara T, Visser F, Luijten PR. MR angiography of the cerebral perforating arteries with magnetization prepared anatomical reference at 7 T: comparison with time-of-flight. J Magn Reson Imaging. 2008;28(6):1519–1526. - PubMed
-
- Rooney WD, Johnson G, Li X, Cohen ER, Kim SG, Ugurbil K, Springer CS., Jr Magnetic field and tissue dependencies of human brain longitudinal 1H2O relaxation in vivo. Magn Reson Med. 2007;57(2):308–318. - PubMed
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources
Research Materials