Complications of pacemaker therapy in adults with congenital heart disease: a multicenter study
- PMID: 23643418
- DOI: 10.1016/j.ijcard.2013.04.114
Complications of pacemaker therapy in adults with congenital heart disease: a multicenter study
Abstract
Background: This study aims to investigate indications and complications of permanent cardiac pacing in adults with congenital heart disease (CHD).
Methods and results: Two-hundred and seventy-four CHD patients were identified who underwent permanent pacemaker implantation between 1972 and 2009. The indication for pacing was acquired sinus node or AV node conduction disease (63%), sinus node or AV node conduction disease after cardiac surgery (28%), and drug/arrhythmia-related indications (9%). Patients with complex CHD received a pacemaker at younger age (23 versus 31 years, p<0.0001) and more often received an epicardial pacing system (51% versus 23%, p<0.0001) compared to those with simple or moderate CHD. Twenty-nine patients (10.6%) had a periprocedural complication during the primary pacemaker implantation (general population: 5.2%). The most common acute complications were lead dysfunction (4.0%), bleeding (2.6%), pocket infection (1.5%) and pneumothorax (1.5%). During a median follow-up of 12 years, pacemaker-related complications requiring intervention occurred in 95 patients (34.6%). The most common late pacemaker-related complications included lead failure (24.8%), pacemaker dysfunction/early battery depletion (5.1%), pacemaker migration (4.7%) and erosion (4.7%). Pacemaker implantation at younger age (<18 years) was an independent predictor of late pacemaker-related complication (adjusted hazard ratio 1.68, 95% confidence interval 1.07 to 2.63, p=0.023).
Conclusions: The risk of periprocedural complications seems higher in the CHD population compared to the general population and more than one-third of CHD patients encountered a pacemaker-related complication during long-term follow-up. This risk increases for those who receive a pacemaker at younger age.
Keywords: Complication; Congenital heart disease; Lead failure; Pacemaker; Tetralogy of Fallot; Transposition of the great arteries.
Crown Copyright © 2013. Published by Elsevier Ireland Ltd. All rights reserved.
Similar articles
-
Atrial-based pacing has no benefit over ventricular pacing in preventing atrial arrhythmias in adults with congenital heart disease.Europace. 2013 Dec;15(12):1757-62. doi: 10.1093/europace/eut213. Epub 2013 Jul 12. Europace. 2013. PMID: 23851513
-
Initial experience of pacing with a lumenless lead system in patients with congenital heart disease.Pacing Clin Electrophysiol. 2009 Nov;32(11):1428-33. doi: 10.1111/j.1540-8159.2009.02487.x. Epub 2009 Sep 21. Pacing Clin Electrophysiol. 2009. PMID: 19769583
-
Pacing in congenital heart disease - A four-decade experience in a single tertiary centre.Int J Cardiol. 2017 Aug 15;241:177-181. doi: 10.1016/j.ijcard.2017.02.151. Epub 2017 Mar 6. Int J Cardiol. 2017. PMID: 28291620
-
Efficacy of prophylactic epicardial pacing leads in children and young adults.Ann Thorac Surg. 2004 Jul;78(1):197-202; discussion 202-3. doi: 10.1016/j.athoracsur.2004.02.008. Ann Thorac Surg. 2004. PMID: 15223427 Review.
-
Indications for pacing in patients with congenital heart disease.Pacing Clin Electrophysiol. 2008 Feb;31 Suppl 1:S17-20. doi: 10.1111/j.1540-8159.2008.00948.x. Pacing Clin Electrophysiol. 2008. PMID: 18226027 Review.
Cited by
-
Comparison of Single- vs. Dual-Lead CIEDs Regarding Perioperative Complications-Analysis of the Quality Assurance Data of the State of North Rhine-Westphalia, Germany.J Clin Med. 2025 May 6;14(9):3203. doi: 10.3390/jcm14093203. J Clin Med. 2025. PMID: 40364234 Free PMC article.
-
Right ventricular pacemaker lead implantation through an occluded vena cava superior baffle after a Senning procedure.HeartRhythm Case Rep. 2025 Feb 10;11(5):412-416. doi: 10.1016/j.hrcr.2025.01.021. eCollection 2025 May. HeartRhythm Case Rep. 2025. PMID: 40656001 Free PMC article. No abstract available.
-
Trends in cardiac implantable electronic device utilization in adults with congenital heart disease: a US nationwide analysis.J Interv Card Electrophysiol. 2024 Mar;67(2):319-328. doi: 10.1007/s10840-023-01601-0. Epub 2023 Jul 1. J Interv Card Electrophysiol. 2024. PMID: 37392273
-
A rat model of complete atrioventricular block recapitulates clinical indices of bradycardia and provides a platform to test disease-modifying therapies.Sci Rep. 2019 May 6;9(1):6930. doi: 10.1038/s41598-019-43300-9. Sci Rep. 2019. PMID: 31061413 Free PMC article.
-
Leadless pacemaker implantation in a patient with a double-chambered right ventricle.BMJ Case Rep. 2022 Jul 15;15(7):e251496. doi: 10.1136/bcr-2022-251496. BMJ Case Rep. 2022. PMID: 35840166 Free PMC article. No abstract available.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical