Comparative accuracy of pulse oximetry and transcutaneous oxygen in assessing arterial saturation in pediatric intensive care
- PMID: 2364712
- DOI: 10.1097/00003246-199007000-00009
Comparative accuracy of pulse oximetry and transcutaneous oxygen in assessing arterial saturation in pediatric intensive care
Abstract
Recent improvements in the noninvasive measurement of oxygen saturation by pulse oximetry (SapO2) have resulted in the development of easily used bedside monitors. We compared the performance of the Biox 3700 and Nellcor N100E models against arterial saturation (SaO2) measured on a Radiometer OSM2 hemoximeter and against transcutaneous SO2 (StcO2) derived from transcutaneous oxygen tensions (PtcO2) (632, Roche Kontron). Twenty-four hemodynamically stable patients were studied, all with indwelling arterial lines, aged between 1 month and 13 yr. Mean differences +/- SD between SapO2 and SaO2 were -0.77 +/- 3.23 for the Nellcor and -2.90 +/- 2.19 for the Biox. The Biox machine generally read lower than the Nellcor with a mean difference of -1.63 +/- 2.65. SaO2 derived from PtcO2 (StcO2) did not differ significantly (-0.27 +/- 2.12) from SaO2. In hemodynamically stable infants and children, StcO2 and SapO2 have similar accuracy limits when compared with each other and with SaO2.
Similar articles
-
Accuracy of two wavelength pulse oximetry in neonates and infants.Pediatr Pulmonol. 1989;6(3):180-2. doi: 10.1002/ppul.1950060310. Pediatr Pulmonol. 1989. PMID: 2470014
-
Pulse oximetry advantages in infants with bronchopulmonary dysplasia.Pediatrics. 1986 Nov;78(5):844-9. Pediatrics. 1986. PMID: 3763297
-
Pulse oximetry in very low birth weight infants with acute and chronic lung disease.Pediatrics. 1987 Apr;79(4):612-7. Pediatrics. 1987. PMID: 2434913
-
[Non-invasive method of arterial oxygen saturation monitoring in the neonatal intensive care unit (with a review of blood oxygen monitoring methods)].Tijdschr Kindergeneeskd. 1988 Feb;56(1):20-6. Tijdschr Kindergeneeskd. 1988. PMID: 3282343 Review. Dutch.
-
Noninvasive monitoring of oxygen and carbon dioxide.Clin Chest Med. 1989 Jun;10(2):239-54. Clin Chest Med. 1989. PMID: 2661121 Review.
Cited by
-
Associations between arterial oxygen saturation, body size and limb measurements among high-altitude Andean children.Am J Hum Biol. 2013 Sep-Oct;25(5):629-36. doi: 10.1002/ajhb.22422. Epub 2013 Aug 1. Am J Hum Biol. 2013. PMID: 23904412 Free PMC article.
-
Comparison of blood gas values in arterial and venous blood.Indian J Pediatr. 2003 Oct;70(10):781-5. doi: 10.1007/BF02723794. Indian J Pediatr. 2003. PMID: 14649471
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources