Marginal and internal fit of zirconia based fixed dental prostheses fabricated with different concepts

Affiliations

Marginal and internal fit of zirconia based fixed dental prostheses fabricated with different concepts

Florian Beuer et al. Clin Cosmet Investig Dent. .

Abstract

The purpose of this in vitro study was to compare the precision of fit of substructures milled from semi-sintered zirconia blocks, fabricated with two different fabrication concepts. Three-unit, posterior fixed dental prostheses (FDP) were fabricated for standardized dies (n = 10) with the laboratory Computer Aided Design (CAD)/Computer Aided Manufacturing (CAM) system Cercon® Brain (Brain) and the centralized CAD/CAM system Compartis Integrated Systems (Compartis). After cementation to the dies, the FDP were embedded and sectioned. Four cross-sections were made of each abutment tooth, and marginal and internal fit were evaluated under an optical microscope. A one-way analysis of variance (ANOVA) was used to compare data (α = 0.05). Mean gap dimensions at the marginal opening for Brain and Compartis were 56.0 (±34.5) μm and 51.7 (±45.2) μm, respectively. Mean internal gap dimensions of 62.8 (±37.5) μm to 164.6 (±33.4) μm were measured depending on the measurement location and the fabrication concept. Mean marginal openings and internal adaptations were not significantly different for both systems. Three out of four measurement locations showed significantly different cement gaps. Within the limitations of this study, the results suggest that the accuracy of both investigated systems is satisfactory for clinical use. The laboratory fabrication exhibited similar accuracy as the centralized manufacturing.

Keywords: CAD; CAM; FDP; all-ceramic; fit; zirconia.

PubMed Disclaimer

Figures

Figure 1
Figure 1
Crown to die diagram showing measurement locations to determine marginal opening (MO; distance between A and B), chamfer area (CA; distance between C and D), axial wall (AW; distance between F and E) and occlusal adaptation (OA; distance between G and H).
Figure 2
Figure 2
Mean gap dimension and standard deviation at marginal opening (MO), chamfer area (CA), axial wall (AW) and occlusal adaptation (OA) measurement locations for both investigated systems.

References

    1. Tinscher J, Natt G, Mohrbotter N, Spiekermann H, Schulze KA. Lifetime of alumina- and zirconia ceramics used for crown and bridge restorations. J Biomed Mater Res B Appl Biomater. 2007;80(2):317–321. - PubMed
    1. Sailer I, Fehér A, Filser F, Gauckler LJ, Lüthy H, Hämmerle CH. Five-year clinical results of zirconia frameworks for posterior fixed partial dentures. Int J Prosthodont. 2007;20(4):383–388. - PubMed
    1. Tinschert J, Natt G, Mautsch W, Augthun M, Spiekermann H. Fracture resistance of lithium disilicate-, alumina-, and zirconia-based three-unit fixed partial dentures: a laboratory study. Int J Prosthodont. 2001;14(3):231–238. - PubMed
    1. Tinschert J, Natt G, Mautsch W, Spiekermann H, Anusavice KJ. Marginal fit of alumina-and zirconia-based fixed partial dentures produced by a CAD/CAM system. Oper Dent. 2001;26(4):367–374. - PubMed
    1. Vult von Steyern P, Carlson P, Nilner K. All-ceramic fixed partial dentures designed according to the DC-Zirkon technique. A 2-year clinical study. J Oral Rehabil. 2005;32(3):180–187. - PubMed