Comparative Study
. 2013 Oct;29(7):1613-8.
doi: 10.1007/s10554-013-0238-2. Epub 2013 May 17.

Coronary CT angiography using the second-generation 320-detector row CT: assessment of image quality and radiation dose in various heart rates compared with the first-generation scanner

Affiliations
Comparative Study

Coronary CT angiography using the second-generation 320-detector row CT: assessment of image quality and radiation dose in various heart rates compared with the first-generation scanner

Nobuo Tomizawa et al. Int J Cardiovasc Imaging. 2013 Oct.

Abstract

To assess the image quality and radiation dose reduction in various heart rates in coronary CT angiography using the second-generation 320-detector row CT compared with the first-generation CT. Ninety-six patients were retrospectively included. The first 48 patients underwent coronary CT angiography with the first-generation 320-detector row CT, while the last 48 patients underwent with the second-generation CT. Subjective image quality was graded using a 4-point scale (4, excellent; 1, unable to evaluate). Image noise and contrast-to-noise ratio were also analyzed. Subgroup analysis was performed based on the heart rate. The mean effective dose was derived from the dose length product multiplied by a conversion coefficient for the chest (κ = 0.014 mSv × mGy(-1) × cm(-1)). The overall subjective image quality score showed no significant difference (3.66 vs 3.69, respectively, p = 0.25). The image quality score of the second-generation group tended to be higher than that of the first-generation group in the 66- to 75-bpm subgroup (3.36 vs 3.53, respectively, p = 0.07). No significant difference was observed in image noise and contrast-to-noise ratio. The overall radiation dose reduced by 24 % (3.3 vs 2.5 mSv, respectively, p = 0.03), and the reduction was substantial in patients with higher heart rate (66- to 75-bpm, 4.3 vs 2.2 mSv, respectively, p = 0.009; >75 bpm, 8.2 vs 3.7 mSv, respectively, p = 0.005). The second-generation 320-detector row CT could maintain the image quality while reducing the radiation dose in coronary CT angiography. The dose reduction was larger in patients with higher heart rate.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Am J Cardiol. 2010 Mar 15;105(6):767-72 - PubMed
    1. Eur Heart J. 2010 Aug;31(15):1908-15 - PubMed
    1. AJR Am J Roentgenol. 2011 Jun;196(6):W729-35 - PubMed
    1. Int J Cardiovasc Imaging. 2013 Jun;29(5):1167-75 - PubMed
    1. Radiology. 2013 Apr;267(1):76-85 - PubMed

Publication types

MeSH terms

LinkOut - more resources