A revision of proton machine quality assurance for wobbled-proton-beam therapy
- PMID: 23689901
- DOI: 10.1007/s12194-013-0217-2
A revision of proton machine quality assurance for wobbled-proton-beam therapy
Abstract
Periodic checks for proton machine quality assurance (QA) are significant for machine users safely and accurately to provide proton-beam treatment for cancer. Our aim in this study was to describe a revision to proton machine QA procedures for wobbled-proton-beam therapy at the Shizuoka Cancer Center (SCC) in Japan. The previous daily, monthly, and annual QA procedures were determined by reference to our past operational experience and to QA papers for medical accelerators. The revised QA procedures were initiated in May 2011 after preliminary measurements to decide baselines for the QA procedures. This paper presents the proton machine QA procedures and the results of representative QA measurements. Three action levels were decided on by reference to the American Association of Physicists in Medicine Task Group 142 report. Tolerances of inspection action were decided on based on the provisional operational results and actual fluctuations of the QA measurement for a year, and those of scheduled action and stop-treatment action were determined by reference to the machine QA papers and those of the inspection action. No deviation from the tolerance of the scheduled action has been observed so far. Although a few QA procedures exceeded the tolerance of the inspection action, these excesses were resolved by inspection and improvement of the respective measuring procedure within the designated QA time. Hereafter, the proton machine QA procedures proposed in this study will be performed continuously at the SCC to assure patient safety and accurate operation of proton therapy.
Similar articles
-
AAPM task group 224: Comprehensive proton therapy machine quality assurance.Med Phys. 2019 Aug;46(8):e678-e705. doi: 10.1002/mp.13622. Epub 2019 Jun 14. Med Phys. 2019. PMID: 31125441
-
Repurposing DailyQA3 for an efficient and spot position sensitive daily quality assurance tool for proton therapy.J Appl Clin Med Phys. 2024 Jul;25(7):e14348. doi: 10.1002/acm2.14348. Epub 2024 Apr 1. J Appl Clin Med Phys. 2024. PMID: 38561975 Free PMC article.
-
A comprehensive and efficient daily quality assurance for PBS proton therapy.Phys Med Biol. 2017 Mar 7;62(5):1661-1675. doi: 10.1088/1361-6560/aa5131. Epub 2017 Feb 6. Phys Med Biol. 2017. PMID: 28166055
-
An overview of the comprehensive proton therapy machine quality assurance procedures implemented at The University of Texas M. D. Anderson Cancer Center Proton Therapy Center-Houston.Med Phys. 2009 Jun;36(6):2269-82. doi: 10.1118/1.3120288. Med Phys. 2009. PMID: 19610316 Review.
-
Machine QA for the Elekta Unity system: A Report from the Elekta MR-linac consortium.Med Phys. 2021 May;48(5):e67-e85. doi: 10.1002/mp.14764. Epub 2021 Apr 1. Med Phys. 2021. PMID: 33577091 Free PMC article. Review.
Cited by
-
Daily QA in proton therapy using a single commercially available detector.J Appl Clin Med Phys. 2014 Nov 8;15(6):5005. doi: 10.1120/jacmp.v15i6.5005. J Appl Clin Med Phys. 2014. PMID: 25493526 Free PMC article.
-
A quality assurance for respiratory gated proton irradiation with range modulation wheel.J Appl Clin Med Phys. 2019 Jan;20(1):258-264. doi: 10.1002/acm2.12526. Epub 2018 Dec 30. J Appl Clin Med Phys. 2019. PMID: 30597762 Free PMC article.
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Research Materials
Miscellaneous