Validation of finite element models for strain analysis of implant-supported prostheses using digital image correlation
- PMID: 23694844
- DOI: 10.1016/j.dental.2013.04.010
Validation of finite element models for strain analysis of implant-supported prostheses using digital image correlation
Abstract
Objectives: A validated numerical model for stress/strain predictions is essential in understanding the biomechanical behavior of implant-supported dental prostheses. The digital image correlation (DIC) method for full-field strain measurement was compared with finite element analysis (FEA) in assessing bone strain induced by implants.
Methods: An epoxy resin model simulating the lower arch was made for the experimental test with acrylic resin replicas of the first premolar and second molar and threaded implants replacing the second premolar and first molar. Splinted (G1/G3) and non-splinted (G2/G4) metal-ceramic screw-retained crowns were fabricated and loaded with (G1/G2) or without (G3/G4) the second molar that provided proximal contact. A single-camera, two-dimensional DIC system was used to record deformation of the resin model surface under a load of 250N. Three-dimensional finite element (FE) models were constructed for the physical models using computer-aided design (CAD) software. Surface strains were used for comparison between the two methods, while internal strains at the implant/resin block interface were calculated using FEA.
Results: Both methods found similar strain distributions over the simulant bone block surface, which indicated possible benefits of having splinted crowns and proximal contact in reducing bone strains. Internal strains predicted by FEA at the implant-resin interface were 8 times higher than those on the surface of the model, and they confirmed the results deduced from the surface strains. FEA gave higher strain values than experiments, probably due to incorrect material properties being used.
Significance: DIC is a useful tool for validating FE models used for the biomechanical analysis of dental prosthesis.
Copyright © 2013 Academy of Dental Materials. Published by Elsevier Ltd. All rights reserved.
Similar articles
-
Comparison of the correlation of photoelasticity and digital imaging to characterize the load transfer of implant-supported restorations.J Prosthet Dent. 2014 Aug;112(2):276-84. doi: 10.1016/j.prosdent.2013.09.029. Epub 2014 Jan 23. J Prosthet Dent. 2014. PMID: 24461947
-
Digital image correlation analysis on the influence of crown material in implant-supported prostheses on bone strain distribution.J Prosthodont Res. 2012 Jan;56(1):25-31. doi: 10.1016/j.jpor.2011.05.003. J Prosthodont Res. 2012. PMID: 21684233
-
Three-dimensional image correlation analyses for strains generated by cement and screw-retained implant prostheses.Clin Implant Dent Relat Res. 2013 Apr;15(2):271-82. doi: 10.1111/j.1708-8208.2011.00411.x. Epub 2011 Dec 15. Clin Implant Dent Relat Res. 2013. PMID: 22171595
-
Biomechanics studies in dentistry: bioengineering applied in oral implantology.J Craniofac Surg. 2009 Jul;20(4):1173-7. doi: 10.1097/SCS.0b013e3181acdb81. J Craniofac Surg. 2009. PMID: 19568186 Review.
-
Stimulating effect of implant loading on surrounding bone. Comparison of three numerical models and validation by in vivo data.Clin Oral Implants Res. 2004 Apr;15(2):239-48. doi: 10.1111/j.1600-0501.2004.01000.x. Clin Oral Implants Res. 2004. PMID: 15008937 Review.
Cited by
-
Voxel-based micro-finite element analysis of dental implants in a human cadaveric mandible: Tissue modulus assignment and sensitivity analyses.J Mech Behav Biomed Mater. 2019 Jun;94:229-237. doi: 10.1016/j.jmbbm.2019.03.008. Epub 2019 Mar 13. J Mech Behav Biomed Mater. 2019. PMID: 30925312 Free PMC article.
-
A novel specimen shape for measurement of linear strain fields by means of digital image correlation.Sci Rep. 2021 Sep 1;11(1):17515. doi: 10.1038/s41598-021-97085-x. Sci Rep. 2021. PMID: 34471200 Free PMC article.
-
Finite element modeling of meniscal tears using continuum damage mechanics and digital image correlation.Sci Rep. 2023 Mar 10;13(1):4039. doi: 10.1038/s41598-023-29111-z. Sci Rep. 2023. PMID: 36899069 Free PMC article.
-
Regulatory Considerations in the Design and Manufacturing of Implantable 3D-Printed Medical Devices.Clin Transl Sci. 2015 Oct;8(5):594-600. doi: 10.1111/cts.12315. Epub 2015 Aug 3. Clin Transl Sci. 2015. PMID: 26243449 Free PMC article. Review.
-
The effect of periapical bone defects on stress distribution in teeth with periapical periodontitis: a finite element analysis.BMC Oral Health. 2023 Dec 8;23(1):980. doi: 10.1186/s12903-023-03546-2. BMC Oral Health. 2023. PMID: 38066540 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous