. 2013:2013:959234.
doi: 10.1155/2013/959234. Epub 2013 Apr 3.

Capsule Endoscopy in Patients with Implantable Electromedical Devices is Safe

Affiliations

Capsule Endoscopy in Patients with Implantable Electromedical Devices is Safe

Lucinda A Harris et al. Gastroenterol Res Pract. 2013.

Abstract

Background and Study Aims. The presence of an implantable electromechanical cardiac device (IED) has long been considered a relative contraindication to the performance of video capsule endoscopy (CE). The primary aim of this study was to evaluate the safety of CE in patients with IEDs. A secondary purpose was to determine whether IEDs have any impact on images captured by CE. Patients and Methods. A retrospective chart review of all patients who had a capsule endoscopy at Mayo Clinic in Scottsdale, AZ, USA, or Rochester, MN, USA, (January 2002 to June 2010) was performed to identify CE studies done on patients with IEDs. One hundred and eighteen capsule studies performed in 108 patients with IEDs were identified and reviewed for demographic data, method of preparation, and study data. Results. The most common indications for CE were obscure gastrointestinal bleeding (77%), anemia (14%), abdominal pain (5%), celiac disease (2%), diarrhea (1%), and Crohn's disease (1%). Postprocedure assessments did not reveal any detectable alteration on the function of the IED. One patient with an ICD had a 25-minute loss of capsule imaging due to recorder defect. Two patients with LVADs had interference with capsule image acquisition. Conclusions. CE did not interfere with IED function, including PM, ICD, and/or LVAD and thus appears safe. Additionally, PM and ICD do not appear to interfere with image acquisition but LVAD may interfere with capsule images and require that capsule leads be positioned as far away as possible from the IED to assure reliable image acquisition.

PubMed Disclaimer

References

    1. Hadithi M, Heine GDN, Jacobs MAJM, V Bodegraven AA, Mulder CJJ. A prospective study comparing video capsule endoscopy with double-balloon enteroscopy in patients with obscure gastrointestinal bleeding. American Journal of Gastroenterology. 2006;101(1):52–57. - PubMed
    1. Appleyard M, Glukhovsky A, Swain P. Wireless-capsule diagnostic endoscopy for recurrent small-bowel bleeding. The New England Journal of Medicine. 2001;344(3):232–233. - PubMed
    1. Neu B, Ell C, May A, et al. Capsule endoscopy versus standard tests in influencing management of obscure digestive bleeding: results from a German multicenter trial. American Journal of Gastroenterology. 2005;100(8):1736–1742. - PubMed
    1. Gay G, Delvaux M, Rey JF. The role of video capsule endoscopy in the diagnosis of digestive diseases: a review of current possibilities. Endoscopy. 2004;36(10):913–920. - PubMed
    1. Niehaus M, Tebbenjohanns J. Electromagnetic interference in patients with implanted pacemakers or cardioverter-defibrillators. Heart. 2001;86(3):246–248. - PMC - PubMed

LinkOut - more resources