Comparative Study
. 2013 Jun;95-B(6):825-30.
doi: 10.1302/0301-620X.95B6.31018.

Inadequate 'three-point' proximal fixation predicts failure of the Gamma nail

Affiliations
Comparative Study

Inadequate 'three-point' proximal fixation predicts failure of the Gamma nail

S G F Abram et al. Bone Joint J. 2013 Jun.

Abstract

The Gamma nail is frequently used in unstable peri-trochanteric hip fractures. We hypothesised that mechanical failure of the Gamma nail was associated with inadequate proximal three-point fixation. We identified a consecutive series of 299 Gamma nails implanted in 299 patients over a five-year period, 223 of whom fulfilled our inclusion criteria for investigation. The series included 61 men and 162 women with a mean age of 81 years (20 to 101). Their fractures were classified according to the Modified Evans' classification and the quality of fracture reduction was graded. The technical adequacy of three points of proximal fixation was recorded from intra-operative fluoroscopic images, and technical inadequacy for each point was defined. All patients were followed to final follow-up and mechanical failures were identified. A multivariate statistical analysis was performed, adjusting for confounders. A total of 16 failures (7.2%) were identified. The position of the lag screw relative to the lateral cortex was the most important point of proximal fixation, and when inadequate the failure rate was 25.8% (eight of 31: odds ratio 7.5 (95% confidence interval 2.5 to 22.7), p < 0.001). Mechanical failure of the Gamma nail in peri-trochanteric femoral fractures is rare (< 1%) when three-point proximal fixation is achieved. However, when proximal fixation is inadequate, failure rates increase. The strongest predictor of failure is positioning the lateral end of the lag screw short of the lateral cortex. Adherence to simple technical points minimises the risk of fixation failure in this vulnerable patient group.

Keywords: Cephalo-medullary nail; Fracture; Gamma nail; Hip fracture; Peri-trochanteric fracture; Trauma.

PubMed Disclaimer

Publication types

MeSH terms