The effect of implant shape and screw pitch on microdamage in mandibular bone
- PMID: 23763648
- DOI: 10.1111/cid.12100
The effect of implant shape and screw pitch on microdamage in mandibular bone
Abstract
Purpose: The aim of this study was to investigate the effect of implant shape and screw pitch on microdamage in bone during insertion of dental implants.
Materials and methods: Thirty custom-made implants (length, 10 mm; diameters, 4.1 mm; cylindrical, tapered, and taper-cylindrical shapes; screw pitches, 1.25 and 0.8 mm; classified as 1.25C, 0.8C, 1.25T, 0.8T, 1.25TC, 0.8TC) were placed with a surgical device in the mandibles of eight goats. Two implant sites were prepared in the edentulous area on each side of the mandible. Implants were placed in a randomized order. Immediately after placement of the implants, the bone blocks with the implants were collected, bulk stained with basic fuchsin, embedded in methyl methacrylate, and sectioned. Histomorphometric quantification of the microcrack length (Cr.Le, μm); microcrack surface density (Cr.Le/B.Ar, μm/mm(2) ), and damaged bone area fraction (DB.Ar/B.Ar, %) were measured.
Results: The Cr.Le, Cr.Le/B.Ar, and DB.Ar/B.Ar values of 0.8TC group were 80.96 ± 17.55, 478.75 ± 51.85, and 4.40 ± 0.36, respectively. All these parameters of microdamage induced by 0.8TC were significantly lower than those induced by other five types of implants (p < .05).
Conclusion: Taper-cylindrical implants with 0.8 mm screw pitch caused the least microdamage to the bone in comparison with the other five types of implants during placement of implants.
Keywords: dental implant; implant shape; mandibular bone; microdamage; screw pitch.
© 2013 Wiley Periodicals, Inc.
Similar articles
-
Microdamage of the cortical bone during mini-implant insertion with self-drilling and self-tapping techniques: a randomized controlled trial.Am J Orthod Dentofacial Orthop. 2012 May;141(5):538-46. doi: 10.1016/j.ajodo.2011.12.016. Am J Orthod Dentofacial Orthop. 2012. PMID: 22554747
-
Influence of surface roughness and shape on microdamage of the osseous surface adjacent to titanium dental implants.Clin Oral Implants Res. 2011 Jun;22(6):613-8. doi: 10.1111/j.1600-0501.2010.02024.x. Epub 2010 Nov 11. Clin Oral Implants Res. 2011. PMID: 21070380
-
A 5-year longitudinal study of the clinical effectiveness of ITI solid-screw implants in the treatment of mandibular edentulism.Int J Oral Maxillofac Implants. 2002 Nov-Dec;17(6):799-810. Int J Oral Maxillofac Implants. 2002. PMID: 12507239 Clinical Trial.
-
Generation of microdamage around endosseous implants.Clin Oral Implants Res. 2009 Dec;20(12):1300-6. doi: 10.1111/j.1600-0501.2009.01808.x. Clin Oral Implants Res. 2009. PMID: 20070741
-
Implant stability measurements for implants placed at the time of extraction: a cohort, prospective clinical trial.J Periodontol. 2005 Mar;76(3):391-7. doi: 10.1902/jop.2005.76.3.391. J Periodontol. 2005. PMID: 15857073 Clinical Trial.
Cited by
-
Repair of microdamage in osteonal cortical bone adjacent to bone screw.PLoS One. 2014 Feb 20;9(2):e89343. doi: 10.1371/journal.pone.0089343. eCollection 2014. PLoS One. 2014. PMID: 24586702 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Research Materials
Miscellaneous