Randomized Controlled Trial
. 2013 Sep;82(9):1571-6.
doi: 10.1016/j.ejrad.2013.04.043. Epub 2013 Jun 15.

Iterative reconstructed ultra high pitch CT pulmonary angiography with cardiac bowtie-shaped filter in the acute setting: effect on dose and image quality

Affiliations
Randomized Controlled Trial

Iterative reconstructed ultra high pitch CT pulmonary angiography with cardiac bowtie-shaped filter in the acute setting: effect on dose and image quality

Steven J Co et al. Eur J Radiol. 2013 Sep.

Abstract

Purpose: To evaluate the effect of a cardiac bowtie-shaped filter in an ultra high pitch CTPA protocol at 100 kV on image quality and radiation dose.

Materials and methods: Retrospective study of 100 patients referred for CTPA. 50 patients scanned with a standard 100 kV protocol at pitch 2.8 (Protocol A) and 50 patients scanned with a 100 kV protocol at pitch 3.2 with a cardiac bowtie-shaped filter (Protocol B). All other scanning parameters kept constant. Images from both groups reconstructed with filtered back projection and iterative reconstruction. Central pulmonary vessel attenuation and background noise were quantitatively measured and signal-to-noise (SNR) and contrast-to-noise (CNR) were calculated. Two radiologists performed qualitative assessment grading visualization of the pulmonary vasculature and noise level. CTDIvol and DLP were recorded and effective dose was calculated.

Results: CTDIvol, DLP and effective dose were significantly (p<0.0001) lower in Protocol B (2.3 ± 0.5 mGy, 78.4 ± 16.5 mGycm, 1.4 ± 0.3 mSy, respectively) compared to Protocol A (4.3 ± 0.5 mGy, 152.0 ± 19.6 mGycm, 2.7 ± 0.3 mSy, respectively). Protocol B had significantly (p<0.0001) higher noise than Protocol A (23.8 ± 6.9 HU vs 36.8 ± 7.3 HU) and lower SNR (11.8 ± 3.7 HU vs 19.2 ± 8.1 HU) and CNR (10.3 ± 3.7 HU vs 24.9 ± 13.4 HU) but there was no significant difference in the subjective visualization of the pulmonary vasculature (p=0.63). Furthermore, iterative reconstruction significantly (p<0.0001) improves image noise (29.4 ± 5.5 HU from 36.8 ± 7.3 HU).

Conclusion: The addition of a cardiac bowtie-shaped filter with an ultra high pitch CTPA protocol at 100 kV resulted in a 48% dose reduction without significantly affecting diagnostic image quality. In addition, the use of iterative reconstruction significantly improves image quality by reducing noise permitting the possibility for further dose reduction strategies.

Keywords: CT; CTPA; Cardiac bowtie-shaped filter; Dose reduction; Pulmonary embolism.