Iterative reconstructed ultra high pitch CT pulmonary angiography with cardiac bowtie-shaped filter in the acute setting: effect on dose and image quality
- PMID: 23777744
- DOI: 10.1016/j.ejrad.2013.04.043
Iterative reconstructed ultra high pitch CT pulmonary angiography with cardiac bowtie-shaped filter in the acute setting: effect on dose and image quality
Abstract
Purpose: To evaluate the effect of a cardiac bowtie-shaped filter in an ultra high pitch CTPA protocol at 100 kV on image quality and radiation dose.
Materials and methods: Retrospective study of 100 patients referred for CTPA. 50 patients scanned with a standard 100 kV protocol at pitch 2.8 (Protocol A) and 50 patients scanned with a 100 kV protocol at pitch 3.2 with a cardiac bowtie-shaped filter (Protocol B). All other scanning parameters kept constant. Images from both groups reconstructed with filtered back projection and iterative reconstruction. Central pulmonary vessel attenuation and background noise were quantitatively measured and signal-to-noise (SNR) and contrast-to-noise (CNR) were calculated. Two radiologists performed qualitative assessment grading visualization of the pulmonary vasculature and noise level. CTDIvol and DLP were recorded and effective dose was calculated.
Results: CTDIvol, DLP and effective dose were significantly (p<0.0001) lower in Protocol B (2.3 ± 0.5 mGy, 78.4 ± 16.5 mGycm, 1.4 ± 0.3 mSy, respectively) compared to Protocol A (4.3 ± 0.5 mGy, 152.0 ± 19.6 mGycm, 2.7 ± 0.3 mSy, respectively). Protocol B had significantly (p<0.0001) higher noise than Protocol A (23.8 ± 6.9 HU vs 36.8 ± 7.3 HU) and lower SNR (11.8 ± 3.7 HU vs 19.2 ± 8.1 HU) and CNR (10.3 ± 3.7 HU vs 24.9 ± 13.4 HU) but there was no significant difference in the subjective visualization of the pulmonary vasculature (p=0.63). Furthermore, iterative reconstruction significantly (p<0.0001) improves image noise (29.4 ± 5.5 HU from 36.8 ± 7.3 HU).
Conclusion: The addition of a cardiac bowtie-shaped filter with an ultra high pitch CTPA protocol at 100 kV resulted in a 48% dose reduction without significantly affecting diagnostic image quality. In addition, the use of iterative reconstruction significantly improves image quality by reducing noise permitting the possibility for further dose reduction strategies.
Keywords: CT; CTPA; Cardiac bowtie-shaped filter; Dose reduction; Pulmonary embolism.
Crown Copyright © 2013. Published by Elsevier Ireland Ltd. All rights reserved.
Similar articles
-
Evaluation of image quality and radiation dose saving comparing knowledge model-based iterative reconstruction on 80-kV CT pulmonary angiography (CTPA) with hybrid iterative reconstruction on 100-kV CT.Emerg Radiol. 2019 Apr;26(2):145-153. doi: 10.1007/s10140-018-1653-4. Epub 2018 Nov 10. Emerg Radiol. 2019. PMID: 30415416
-
Clinical utility of ultra high pitch dual source thoracic CT imaging of acute pulmonary embolism in the emergency department: are we one step closer towards a non-gated triple rule out?Eur J Radiol. 2013 Oct;82(10):1793-8. doi: 10.1016/j.ejrad.2013.05.003. Epub 2013 Jun 3. Eur J Radiol. 2013. PMID: 23743054
-
Image quality and radiation exposure at pulmonary CT angiography with 100- or 120-kVp protocol: prospective randomized study.Radiology. 2007 Nov;245(2):577-83. doi: 10.1148/radiol.2452061919. Radiology. 2007. PMID: 17940308 Clinical Trial.
-
Iterative reconstruction of pulmonary MDCT angiography: effects on image quality, effective dose and estimated organ dose to the breast.Biomed Pap Med Fac Univ Palacky Olomouc Czech Repub. 2014 Jun;158(2):259-64. doi: 10.5507/bp.2013.059. Epub 2013 Sep 12. Biomed Pap Med Fac Univ Palacky Olomouc Czech Repub. 2014. PMID: 24026145 Review.
-
Pulmonary CT angiography as first-line imaging for PE: image quality and radiation dose considerations.AJR Am J Roentgenol. 2013 Mar;200(3):522-8. doi: 10.2214/AJR.12.9928. AJR Am J Roentgenol. 2013. PMID: 23436840 Review.
Cited by
-
CT angiography in the diagnosis of cardiovascular disease: a transformation in cardiovascular CT practice.Quant Imaging Med Surg. 2014 Oct;4(5):376-96. doi: 10.3978/j.issn.2223-4292.2014.10.02. Quant Imaging Med Surg. 2014. PMID: 25392823 Free PMC article. Review.
-
High-Pitch CT Pulmonary Angiography in Third Generation Dual-Source CT: Image Quality in an Unselected Patient Population.PLoS One. 2016 Feb 12;11(2):e0146949. doi: 10.1371/journal.pone.0146949. eCollection 2016. PLoS One. 2016. PMID: 26872262 Free PMC article.
-
High-pitch computed tomography pulmonary angiography with iterative reconstruction at 80 kVp and 20 mL contrast agent volume.Eur Radiol. 2014 Dec;24(12):3260-8. doi: 10.1007/s00330-014-3365-9. Epub 2014 Aug 7. Eur Radiol. 2014. PMID: 25100336 Clinical Trial.
-
High pitch, low voltage dual source CT pulmonary angiography: assessment of image quality and diagnostic acceptability with hybrid iterative reconstruction.Emerg Radiol. 2015 Apr;22(2):117-23. doi: 10.1007/s10140-014-1230-4. Epub 2014 Jul 4. Emerg Radiol. 2015. PMID: 24993583
-
Sparse sampling computed tomography (SpSCT) for detection of pulmonary embolism: a feasibility study.Eur Radiol. 2019 Nov;29(11):5950-5960. doi: 10.1007/s00330-019-06217-5. Epub 2019 May 9. Eur Radiol. 2019. PMID: 31073861
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical