Quantification of tissue shrinkage and dehydration caused by microwave ablation: experimental study in kidneys for the estimation of effective coagulation volume
- PMID: 23792128
- DOI: 10.1016/j.jvir.2013.04.008
Quantification of tissue shrinkage and dehydration caused by microwave ablation: experimental study in kidneys for the estimation of effective coagulation volume
Abstract
Purpose: To quantify the extent of tissue shrinkage and dehydration caused by microwave (MW) ablation in kidneys for estimation of effective coagulation volume.
Materials and methods: MW ablations were carried out in ex vivo porcine kidneys. Six study groups were defined: groups 1A, 2A, and 3A for MW ablation (90 W for 5 min, 7.5 min, or 10 min), and groups 1B, 2B, and 3B for control (without MW ablation). Pre- and postinterventional volume analyses were performed. Effective coagulation volumes (original tissue included in coagulation) were determined. Postinterventional dehydration analyses were performed with calculation of mean mass fractions of water.
Results: Mean deployed energies were 21.6 kJ ± 1.1 for group 1A, 29.9 kJ ± 1.0 for group 2A, and 42.1 kJ ± 0.5 kJ for group 3A, and were significantly different (P < .0001). Differences between pre- and postinterventional volumes were -3.8% ± 0.6 for group 1A, -5.6% ± 0.9 for group 2A, and -7.2% ± 0.4 for group 3A, and -1.1% ± 0.3 for group 1B, -1.8% ± 0.4 for group 2B, and -1.1% ± 0.4 for group 3B. Postinterventional volumes were significantly smaller than preinterventional volumes for all groups (P < .01). Underestimations of effective coagulation volume from visualized coagulation volume were 26.1% ± 3.5 for group 1A, 35.2% ± 11.2 for group 2A, and 42.1% ± 4.9 for group 3A, which were significantly different (P < .01). Mean mass fractions of water were 64.2% ± 1.4 for group 1A, 63.2% ± 1.7 for group 2A, and 62.6% ± 1.8% for group 3A, with significant differences versus corresponding control groups (P < .01).
Conclusions: For MW ablation in kidneys, underestimation of effective coagulation volume based on visualized coagulation volume is significantly greater with greater deployed energy. Therefore, local dehydration with tissue shrinkage is a potential contributor.
Keywords: AKS; MW; RF; appearance of kidney surface; microwave; radiofrequency.
Copyright © 2013 SIR. Published by Elsevier Inc. All rights reserved.