Comparative Study
. 2013 Oct;10(10):1444-50.
doi: 10.1016/j.hrthm.2013.06.021. Epub 2013 Jun 28.

Transvenous extraction of implantable cardioverter-defibrillator leads under advisory--a comparison of Riata, Sprint Fidelis, and non-recalled implantable cardioverter-defibrillator leads

Affiliations
Comparative Study

Transvenous extraction of implantable cardioverter-defibrillator leads under advisory--a comparison of Riata, Sprint Fidelis, and non-recalled implantable cardioverter-defibrillator leads

Michael P Brunner et al. Heart Rhythm. 2013 Oct.

Abstract

Background: Comparative safety and efficacy associated with transvenous lead extraction (TLE) of recalled and non-recalled implantable cardioverter-defibrillator (ICD) leads has not been well characterized.

Objectives: To compare the indications, techniques, and procedural outcomes of recalled vs non-recalled ICD lead extraction procedures.

Methods: TLE procedures performed at our institution from June 2002 to June 2012 in which Riata, Sprint Fidelis, or non-recalled ICD leads were extracted were included in the analysis.

Results: ICD lead extraction procedures were performed in 1079 patients, including 430 patients with recalled leads (121 Riata, 308 Sprint Fidelis, and 1 Riata and Sprint Fidelis) and 649 patients with non-recalled ICD leads. A total of 2056 chronic endovascular leads were extracted, of which 1215 (59.1%) were ICD leads. Overall, there was 96.8% complete procedural success, 99.1% clinical success, and 0.9% failure, with 3.9% minor complications and 1.5% major complications. Procedural outcomes for Riata and Sprint Fidelis TLE procedures were no different. Lead implant duration was significantly less in recalled than in non-recalled ICD lead TLE procedures. Complete procedural success was higher in recalled (424 of 430 [98.6%]) than in non-recalled (621 of 649 [95.7%]; P = .007) ICD lead TLE procedures. Minor complications were lower in recalled (10 of 430 [2.3%]) than in non-recalled (32 of 649 [5.0%]; P = .030) ICD lead TLE procedures. Rates of clinical success, failure, and major complications were no different in the recalled and non-recalled ICD lead TLE procedures.

Conclusion: In our experience, recalled ICD leads were extracted with safety and efficacy comparable to that of non-recalled ICD leads.

Keywords: Advisory; EC; Extraction; FDA; Food and Drug Administration; HRS; Heart Rhythm Society; ICD; Implantable cardioverter-defibrillator; Lead; SVC; TLE; externalized conductor; implantable cardioverter-defibrillator; superior vena cava; transvenous lead extraction.

PubMed Disclaimer