Fracture of ceramic bearing surfaces following total hip replacement: a systematic review

Francesco Traina et al. Biomed Res Int. 2013.

Abstract

Ceramic bearing surfaces are increasingly used for total hip replacement, notwithstanding that concern is still related to ceramic brittleness and its possible mechanical failure. The aim of this systematic review is to answer three questions: (1) Are there risk factors for ceramic component fracture following total hip replacement? (2) Is it possible to perform an early diagnosis of ceramic component failure before catastrophic fracture occurs? (3) Is it possible to draw guidelines for revision surgery after ceramic components failure? A PubMed and Google Scholar search was performed and reference citations from publications identified in the literature search were reviewed. The use of 28 mm short-neck femoral head carries an increased risk of fracture. Acetabular component malposition might increase the risk of ceramic liner fractures. Synovial fluid microanalysis and CT scan are promising in early diagnosis of ceramic head and liner failure. Early revision is suggested in case of component failure; no consensus exists about the better coupling for revision surgery. Ceramic brittleness remains a major concern. Due to the increased number of ceramic on ceramic implants, more revision surgeries and reports on ceramic components failure are expected in the future. An algorithm of diagnosis and treatment for ceramic hip failure is proposed.

PubMed Disclaimer

Figures

Figure 1
Figure 1
(a) Preoperative radiograph showing catastrophic failure of a ceramic liner. (b) The retrieved ceramic head grossly damaged because of contact with metal back. (c) The retrieved ceramic liner. (d) Postoperative radiograph after cup revision and bearing surfaces exchange to metal on metal coupling.
Figure 2
Figure 2
Diagram showing the process of manuscripts selection.
Figure 3
Figure 3
Algorithm with guidelines for the diagnosis and treatment of fractured ceramic bearings following THR.

Similar articles

Cited by

References

    1. Bos I, Willmann G. Morphologic characteristics of periprosthetic tissues from hip prostheses with ceramic-ceramic couples: a comparative histologic investigation of 18 revision and 30 autopsy cases. Acta Orthopaedica Scandinavica. 2001;72(4):335–342. - PubMed
    1. Christel PS. Biocompatibility of surgical-grade dense polycrystalline alumina. Clinical Orthopaedics and Related Research. 1992;(282):10–18. - PubMed
    1. García-Rey E, García-Cimbrelo E. Polyethylene in total hip arthroplasty: half a century in the limelight. Journal of Orthopaedics and Traumatology. 2010;11(2):67–72. - PMC - PubMed
    1. Gudena R, Kuna S, Pradhan N. Aseptic loosening of total hip replacement presenting as an anterior thigh mass. Musculoskeletal Surgery. 2011 - PubMed
    1. Spinarelli A, Patella V, Petrera M, Abate A, Pesce V, Patella S. Heterotopic ossification after total hip arthroplasty: our experience. Musculoskeletal surgery. 2011;95(1):1–5. - PubMed