Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Case Reports
. 2013 Jul 5:6:29-32.
doi: 10.2147/IMCRJ.S45784. Print 2013.

Sinus venosus atrial septal defect: a rare cause of misplacement of pacemaker leads

Affiliations
Case Reports

Sinus venosus atrial septal defect: a rare cause of misplacement of pacemaker leads

Malick Bodian et al. Int Med Case Rep J. .

Abstract

Routine implantation of pacemakers and implantable cardioverter defibrillators is not commonly associated with complications. However, in some cases we see misplacement of pacemaker leads which is most often related to the presence of underlying cardiac anomalies. We report the case of misplacement of a pacemaker lead into the left ventricle of a 56-year-old patient paced in VVI/R mode and with a tined type pacemaker lead because of a symptomatic complete atrioventricular block. Electrocardiogram showed a pacemaker-generated rhythm with a right bundle branch block pattern. Chest X-ray showed the pacemaker lead located relatively high in relation to the diaphragm. Echocardiography visualized the pacemaker lead in the left heart chambers (atrium and ventricle), hence confirming its aberrant course. Further, the defect causing its passage to the left heart chambers was a sinus venosus atrial septal defect. The patient reported no complication related to the misplacement of the lead. After a brief period of oral anticoagulation, the lead was inserted into the right ventricle by percutaneous technique.

Keywords: lead misplacement; pacemaker; sinus venosus atrial septal defect.

PubMed Disclaimer

Figures

Figure 1
Figure 1
(A) Post-implantation electrocardiogram showing ventricular spikes and a right bundle branch block pattern. (B) Post-implantation chest X-ray showing the pacemaker lead (arrow) located relatively high in relation to the diaphragm.
Figure 2
Figure 2
(A and B) Two-dimensional transthoracic echocardiography with a parasternal long axis view showing pacemaker lead (arrow) in the left heart chambers. Abbreviations: LA, left atrium; LV, left ventricle; RV, right ventricle.
Figure 3
Figure 3
(A and B) Two-dimensional echocardiographic subcostal view showing the sinus venosus atrial septal defect (defect in the superior-posterior portion of the interatrial septum at the junction of the superior vena cava and right atrium). Abbreviations: ASD, atrial septal defect; LA, left atrium; RA, right atrium.
Figure 4
Figure 4
Post reimplantation electrocardiogram showing a left bundle branch block pattern, indicating the presence of the lead in the right cavities.

References

    1. Vanhercke D, Heytens W, Verloove H. Eight years of left ventricle pacing due to inadvertent malposition of a transvenous pacemaker lead in the left ventricle. Eur J Echocardiogr. 2008;9:825–827. - PubMed
    1. Tobin AM, Grodman RS, Fisherkeller M, Nicolosi R. Two-dimensional echocardiographic localization of a malpositioned pacing catheter. Pacing Clin Electrophysiol. 1983;6:291–299. - PubMed
    1. Gondi B, Nanda NC. Real-time two-dimensional echocardiography features of pacemaker perforation. Circulation. 1981;64:97–106. - PubMed
    1. Winner SJ, Boon NA. Transvenous pacemaker electrodes placed unintentionally in the left ventricle: three cases. Postgrad Med J. 1989;65:98–102. - PMC - PubMed
    1. Mazzetti H, Dussaut A, Tentori C, Dussaut E, Lazzari JO. Transarterial permanent pacing of the left ventricle. Pacing Clin Electrophysiol. 1990;13:588–592. - PubMed

Publication types

LinkOut - more resources