Effect of pupil size on optical quality parameters in astigmatic eyes using a double-pass instrument
- PMID: 23865038
- PMCID: PMC3706062
- DOI: 10.1155/2013/124327
Effect of pupil size on optical quality parameters in astigmatic eyes using a double-pass instrument
Abstract
Purpose: To objectively determine the effects of pupil size on optical quality parameters in astigmatic eyes using a double-pass instrument.
Methods: We examined twenty-two eyes of 22 healthy volunteers (mean age ± standard deviation, 27.1 ± 2.8 years) who had no ophthalmic diseases other than refractive errors (manifest cylinder ≤ 0.25 diopters (D)). After we fully corrected cycloplegic refraction, we created with-the-rule astigmatism of 1, 2, and 3 diopters in these eyes and then quantitatively assessed the modulation transfer function (MTF) cutoff frequency and the Strehl2D ratio with 2-, 4-, and 6-mm pupil sizes using the Optical Quality Analysis System.
Results: The MTF cutoff frequency and the Strehl2D ratio decreased significantly as the amount of astigmatism increased at each pupil size (P < 0.001 for 2, 4, and 6 mm, analysis of variance). They also decreased significantly with an increase in pupil size at each diopter of astigmatism (P < 0.001 for 0, 1, 2, and 3 D). Multiple comparisons demonstrated a significant difference between measurements made for a 2-mm pupil and for a 6-mm pupil at each diopter of astigmatism (P < 0.001 for 0, 1, 2, and 3 D, Dunnett test) and those made for a 4-mm pupil and for a 6-mm pupil at each diopter of astigmatism (P < 0.001 for 0 D, P < 0.05 1, 2, and 3 D).
Conclusions: Eyes with larger pupils had lower optical quality even when they were astigmatic. It may be necessary to correct the preexisting astigmatism to acquire excellent visual performance, especially in astigmatic eyes with larger pupils.
Figures



References
-
- Read SA, Collins MJ, Carney LG. A review of astigmatism and its possible genesis: invited review. Clinical and Experimental Optometry. 2007;90(1):5–19. - PubMed
-
- Ferrer-Blasco T, Montés-Micó R, Peixoto-de-Matos SC, González-Méijome JM, Cerviño A. Prevalence of corneal astigmatism before cataract surgery. Journal of Cataract & Refractive Surgery. 2009;3:70–75. - PubMed
-
- Hoffmann PC, Hütz WW. Analysis of biometry and prevalence data for corneal astigmatism in 23 239 eyes. Journal of Cataract and Refractive Surgery. 2010;36(9):1479–1485. - PubMed
-
- Güell JL, Pujol J, Arjona M, Diaz-Douton F, Artal P. Optical Quality Analysis System: instrument for objective clinical evaluation of ocular optical quality. Journal of Cataract and Refractive Surgery. 2004;30(7):1598–1599. - PubMed
-
- Vilaseca M, Padilla A, Pujol J, Ondategui JC, Artal P, Güell JL. Optical quality one month after Verisyse and Veriflex phakic IOL implantation and Zeiss MEL 80 LASIK for myopia from 5.00 to 16.50 diopters. Journal of Refractive Surgery. 2009;25(8):689–698. - PubMed
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous