[Biomechanical effects of iliac screw plates on stability of lumbo-iliac fixation construct]
- PMID: 23879102
[Biomechanical effects of iliac screw plates on stability of lumbo-iliac fixation construct]
Abstract
Objective: To evaluate the biomechanical effect of a self-made iliac screw plate on the stability of lumbo-iliac fixation construct before and after fatigue loading.
Methods: Twelve fresh lumbo-pelvic specimens from donated adult cadavers with formalin embalm were used in the study. According to whether use the iliac screw plate or not, the specimens were randomly assigned into group A (with iliac screw plate, n=6) and group B (without iliac screw plate, n=6). The bone mineral density (BMD) of L(t-4) was measured using dual-energy radiograph absorptiometry. The pedicle screw and iliac screw fixation were given at L3-5, and bilateral facetectomy and diskectomy at L5, S1 level were performed to prepare the model of the intervertebral destabilization. The biomechanical testing was conducted on a material testing machine under 0-600 N compression and -7-7 N.m torsion loading modes for the initial compressive stiffness and torsional stiffness evaluation. And then 20 000 cyclic compressive loading of 40-400 N was given to the specimen, the stiffness evaluation was repeated. Then the maximum pull-out strength of screws at every level was measured and compared. Gross observation and radiological observation were performed during experiment.
Results: The BMD values of groups A and B were (1.15 +/-0.13) g/cm(2) and (1.12 +/-0.11) g/cm(2) respectively, showing no significant difference between 2 groups (t=0.428, P=0.678). All pedicle screws and iliac screws were inserted in good position; no loosening or breaking of screw was observed during loading. After fatigue loading, the incidence of halo ring around the iliac screws of groups A and B was 16.7% (1/6) and 50.0% (3/6), respectively. The compressive stiffness and torsional stiffness after fatigue loading were significantly lower than those in initial state in groups A and B (P < 0.05); there was no significant difference in compressive stiffness and torsional stiffness between groups A and B before fatigue loading (P > 0.05). However, group A had higher compressive stiffness than group B (t=2.664, P=0.024) after fatigue loading, and there was no significant difference in torsional stiffness between 2 groups (t=0.410, P=0.690). No significant difference was found in screw pull-out strength of pedicle screws at L3, L4, and L5 levels between groups A and B (P > 0.05); however, the pull-out strength of the iliac screws in group A was significantly higher than that in group B (t=3.398, P=0.007). In groups A and B, the pull-out strength of L3 screw was significantly lower than that of L4 and L5 screws (P < 0.05). In group A, pull-out strength of the iliac screws was significantly higher than that of L3, L4, and L5 screws (P < 0.05); in group B, the pull-out strength of iliac screws was significantly lower than that of L4 and L5 screws (P < 0.05).
Conclusion: In the lumbo-iliac reconstruction, the use of iliac screw plate could resist iliac screw loosening, therefore, it has the potential to increase the stability oflumbo-iliac fixation construct.
Similar articles
-
Effect of iliac screw insertion depth on the stability and strength of lumbo-iliac fixation constructs: an anatomical and biomechanical study.Spine (Phila Pa 1976). 2009 Jul 15;34(16):E565-72. doi: 10.1097/BRS.0b013e3181ac8fc4. Spine (Phila Pa 1976). 2009. PMID: 19770599
-
Effect of the degree of osteoporosis on the biomechanical anchoring strength of the sacral pedicle screws: an in vitro comparison between unaugmented bicortical screws and polymethylmethacrylate augmented unicortical screws.Spine (Phila Pa 1976). 2010 Sep 1;35(19):E925-31. doi: 10.1097/BRS.0b013e3181c5fb21. Spine (Phila Pa 1976). 2010. PMID: 20098349
-
Biomechanical advantages of dual over single iliac screws in lumbo-iliac fixation construct.Eur Spine J. 2010 Jul;19(7):1121-8. doi: 10.1007/s00586-010-1343-8. Epub 2010 Mar 2. Eur Spine J. 2010. PMID: 20195650 Free PMC article.
-
Dual iliac screws in spinopelvic fixation: a systematic review.Eur Spine J. 2019 Sep;28(9):2053-2059. doi: 10.1007/s00586-019-06065-3. Epub 2019 Jul 12. Eur Spine J. 2019. PMID: 31300882
-
[Application of second sacral alar-iliac screw technique for reconstruction of spinopelvic stability].Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi. 2018 Jun 15;32(6):764-768. doi: 10.7507/1002-1892.201711124. Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi. 2018. PMID: 29905058 Free PMC article. Review. Chinese.