Abstract
The aim of this study was to compare the performance of four methods that are widely used in the clinical microbiology laboratory for identification of Pasteurella species. The 4 methods evaluated were VITEK2, VITEK MS (BioMerieux), and Bruker Biotyper MS (Bruker) as well as traditional biochemical tests. Sequencing of the sodA gene was used as the reference method. Sixty-five isolates of Pasteurella spp. from 65 patients were analyzed. One Pasteurella multocida isolate from American Type Culture Collection (Manassas, VA, USA) was used as a reference. Traditional biochemical tests accurately identified 62/66 (94%) isolates. Both Bruker and Vitek matrix-assisted laser desorption ionization-time of flight (MALDI-TOF) identified 59/66 (89%) strains, but VITEK2 could only identify 32/66 (48.5%) isolates correctly. The mean time to identification using biochemical tests was 20 hours; VITEK2 took 6 hours and MALDI-TOF approximately 10 minutes. In conclusion, MALDI-TOF is a quick method, which accurately identified most isolates of Pasteurella to the species level. Thus, MALDI-TOF constitutes a valuable diagnostic tool in the clinical laboratory.
Keywords: MALDI-TOF; Pasteurella spp; Vitek2; animal bite infections.