Review
. 2013 Oct;58(10):1694-703.
doi: 10.4187/respcare.02786. Epub 2013 Jul 30.

AARC clinical practice guideline: blood gas analysis and hemoximetry: 2013

Affiliations
Review

AARC clinical practice guideline: blood gas analysis and hemoximetry: 2013

Michael D Davis et al. Respir Care. 2013 Oct.

Abstract

We searched MEDLINE, CINAHL, and Cochrane Library database for articles published between January 1990 and December 2012. The update of this clinical practice guideline is based on 237 clinical trials, 54 reviews, and 23 meta-analyses on blood gas analysis (BGA) and hemoximetry. The following recommendations are made following the Grading of Recommendations Assessment, Development, and Evaluation scoring system. BGA and hemoximetry are recommended for evaluating a patient's ventilatory, acid-base, and/or oxygenation status. BGA and hemoximetry are suggested for evaluating a patient's response to therapeutic interventions. BGA and hemoximetry are recommended for monitoring severity and progression of documented cardiopulmonary disease processes. Hemoximetry is recommended to determine the impact of dyshemoglobins on oxygenation. Capillary BGA is not recommended to determine oxygenation status. Central venous BGA and hemoximetry are suggested to determine oxygen consumption in the setting of early goal-directed therapies. For the assessment of oxygenation, a peripheral venous P(O2) is not recommended as a substitute for an arterial blood measurement (P(aO2)). It is not recommended to use venous P(CO2) and pH as a substitute for arterial blood measurement of P(aCO2) and pH. It is suggested that hemoximetry is used in the detection and evaluation of shunts during diagnostic cardiac catheterization.

Keywords: blood gas analysis; blood gases; guidelines; hemoximetry.