Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Clinical Trial
. 2013 Jul 1;7(4):842-53.
doi: 10.1177/193229681300700406.

Performance evaluation of three continuous glucose monitoring systems: comparison of six sensors per subject in parallel

Affiliations
Clinical Trial

Performance evaluation of three continuous glucose monitoring systems: comparison of six sensors per subject in parallel

Guido Freckmann et al. J Diabetes Sci Technol. .

Abstract

Background: This study is aimed at comparing the performance of three continuous glucose monitoring (CGM) systems following the Clinical and Laboratory Standards Institute's POCT05-A guideline, which provides recommendations for performance evaluation of CGM systems.

Methods: A total of 12 subjects with type 1 diabetes were enrolled in this study. Each subject wore six CGM systems in parallel, two sensors of each CGM system [FreeStyle Navigator™ (Navigator), MiniMed Guardian® REAL-Time with Enlite sensor (Guardian), DexCom™ Seven® Plus 3rd generation (Seven Plus)]. Each sensor was used for the lifetime specified by the manufacturer. To follow POCT05-A recommendations, glucose excursions were induced on two separate occasions, and venous and capillary blood glucose (BG) concentrations were obtained every 15 min for five consecutive hours. Capillary BG concentrations were measured at least once per hour during the day and once at night. Parameters investigated were CGM-to-BG differences [mean absolute relative difference (MARD)] and sensor-to-sensor differences [precision absolute relative difference (PARD)].

Results: Compared with capillary BG reference readings, the Navigator showed the lowest MARD, with 12.1% overall and 24.6% in the hypoglycemic range; for the Guardian and the Seven Plus, MARD was 16.2%/34.9% and 16.3%/32.7%, respectively. PARD also was lowest for the Navigator (9.6%/9.8%), followed by the Seven Plus (16.7%/25.5%) and the Guardian (18.1%/20.2%). During induced glucose excursions, MARD between CGM and BG was, again, lowest for the Navigator (14.3%), followed by the Seven Plus (15.8%) and the Guardian (19.2%).

Conclusions: In this study, two sensors of each of the three CGM systems were compared in a setting following POCT05-A recommendations. The Navigator CGM system achieved more accurate results than the Guardian or the Seven Plus with respect to MARD and PARD. Performance in the hypoglycemic range was markedly worse for all CGM systems when compared with BG results.

PubMed Disclaimer

Figures

Figure 1.
Figure 1.
Study procedures as displayed by a representative experiment. The red, solid rectangle marks the core phase. The induced glucose excursions are highlighted and enlarged in red, dotted rectangles. cap, capillary; CHO, carbohydrate.
Figure 2.
Figure 2.
Precision absolute relative difference over study days. For the Navigator and the Seven Plus, PARD decreases over time, i.e., precision increases.
Figure 3. (A)
Figure 3. (A)
Difference plot with ISO 15197:2003 system accuracy limits (±15 mg/dl for reference BG values <75 mg/dl and ±20% for reference BG values ≥75 mg/dl) for Navigator CGM values compared with Navigator-BG. (B) Difference plot with ISO 15197:2003 system accuracy limits for Guardian CGM values compared with Navigator-BG. (C) Difference plot with ISO 15197:2003 system accuracy limits for Seven Plus CGM values compared with Navigator-BG.
Figure 4.
Figure 4.
Cumulative percentage of CGM values differing <5%, 10%, 20%, and 30% from the capillary reference values. Results displayed for core phase (days 2 to 5): overall, <70, 70 to 180, and >180 mg/dl.
Figure 5.
Figure 5.
Relative number of occurrence of capillary and venous glucose rates of change (obtained from Navigator-BG and YSI 2300, respectively) during induced glucose excursions. Measurements were performed every 15 min.

References

    1. Clinical and Laboratory Standards Institute Performance metrics for continuous interstitial glucose monitoring; approved guideline. http://shopping.netsuite.com/c.1253739/site/Sample_pdf/POCT05A__sample.pdf . Accessed June 14, 2013.
    1. Clarke WL, Anderson S, Farhy L, Breton M, Gonder-Frederick L, Cox D, Kovatchev B. Evaluating the clinical accuracy of two continuous glucose sensors using continuous glucose-error grid analysis. Diabetes Care. 2005;28(10):2412–2417. - PubMed
    1. Weinstein RL, Schwartz SL, Brazg RL, Bugler JR, Peyser TA, McGarraugh GV. Accuracy of the 5-day FreeStyle Navigator Continuous Glucose Monitoring System: comparison with frequent laboratory reference measurements. Diabetes Care. 2007;30(5):1125–1130. - PubMed
    1. Wilson DM, Beck RW, Tamborlane WV, Dontchev MJ, Kollman C, Chase P, Fox LA, Ruedy KJ, Tsalikian E, Weinzimer SA, DirecNet Study Group The accuracy of the FreeStyle Navigator continuous glucose monitoring system in children with type 1 diabetes. Diabetes Care. 2007;30(1):59–64. - PMC - PubMed
    1. Mastrototaro J, Shin J, Marcus A, Sulur G. STAR 1 Clinical Trial Investigators. The accuracy and efficacy of real-time continuous glucose monitoring sensor in patients with type 1 diabetes. Diabetes Technol Ther. 2008;10(5):385–390. - PubMed

Publication types